Your browser doesn't support javascript.
loading
A comprehensive LC-MS/MS method for the simultaneous measurement of 24 adrenal steroids: From research to clinical practice.
Yu, Songlin; Yin, Yicong; Zou, Yutong; Li, Qianqian; Yu, Jialei; Xie, Shaowei; Luo, Wei; Ma, Xiaoli; Wang, Danchen; Lin, Liling; Qiu, Ling.
Afiliación
  • Yu S; Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing 100730, China.
  • Yin Y; Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing 100730, China.
  • Zou Y; Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing 100730, China.
  • Li Q; Waters Technologies (Shanghai) Co., Ltd., Shanghai, China.
  • Yu J; Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing 100730, China.
  • Xie S; Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing 100730, China.
  • Luo W; Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing 100730, China.
  • Ma X; Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing 100730, China.
  • Wang D; Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing 100730, China.
  • Lin L; Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing 100730, China.
  • Qiu L; Department of Laboratory Medicine, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Sciences, Beijing 100730, China; Severe and Rare Diseases, Peking Union Medical College Hospital, Peking Union Medical College & Chinese Academy of Medical Scie
Article en En | MEDLINE | ID: mdl-38103306
ABSTRACT
Steroids are essential in the differential diagnosis of congenital adrenal hyperplasia (CAH) subtypes; however, they may confuse physicians with multifarious results. In this study, we established a liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the simultaneous measurement of 24 steroids and developed a steroid metabolite pathway-based report to aid physicians in understanding these results. Solid-phase extraction was used to concentrate and purify target plasma steroids. The linearity, precision, recovery, and matrix effects were thoroughly evaluated. PowerBuilder was used to transfer the results from LC-MS/MS to the graphic report in a laboratory information management system (LIS) and was applied to different subtypes of CAH. Twenty-four steroids were separated and analyzed in one sample preparation and two injections using LC-MS/MS. The linearity of the steroids was excellent, with coefficients of linear regression greater than 0.99. The relative recovery ranged from 90.0 to 107.1 %, whereas the intra- and total coefficient variations were 1.6 âˆ¼ 8.7 % and 2.0 âˆ¼ 9.9 %, respectively. Matrix effects were compensated after internal standard correction. A graphic combination report mode was established and used to effectively identify CAH subtypes. In conclusion, a useful LC-MS/MS method and graphic combination report of 24 steroids based on their metabolite pathways were established.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Hiperplasia Suprarrenal Congénita / Espectrometría de Masas en Tándem Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci / J. chromatogr. B / Journal of chromatography. B (Print) Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2024 Tipo del documento: Article

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Hiperplasia Suprarrenal Congénita / Espectrometría de Masas en Tándem Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci / J. chromatogr. B / Journal of chromatography. B (Print) Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2024 Tipo del documento: Article