Mitigation of electromagnetic pulses interfering with Thomson parabola ion spectrometers at XG-III laser facility.
Rev Sci Instrum
; 95(1)2024 Jan 01.
Article
en En
| MEDLINE
| ID: mdl-38197767
ABSTRACT
The Thomson parabola ion spectrometer is vulnerable to intense electromagnetic pulses (EMPs) generated by a high-power laser interacting with solid targets. A metal shielding cage with a circular aperture of 1 mm diameter is designed to mitigate EMPs induced by a picosecond laser irradiating a copper target in an experiment where additionally an 8-ns delayed nanosecond laser is incident into an aluminum target at the XG-III laser facility. The implementation of the shielding cage reduces the maximum EMP amplitude inside the cage to 5.2 kV/m, and the simulation results indicate that the cage effectively shields electromagnetic waves. However, the laser-accelerated relativistic electrons which escaped the target potential accumulate charge on the surface of the cage, which is responsible for the detected EMPs within the cage. To further alleviate EMPs, a lead wall and an absorbing material (ECCOSORB AN-94) were added before the cage, significantly blocking the propagation of electrons. These findings provide valuable insights into EMP generation in large-scale laser infrastructures and serve as a foundation for electromagnetic shielding design.
Texto completo:
1
Base de datos:
MEDLINE
Idioma:
En
Revista:
Rev Sci Instrum
Año:
2024
Tipo del documento:
Article