Dislocation force of scleral flange-fixated intraocular lens haptics.
Stunf Pukl, Spela; Kronschläger, Martin; Ruiss, Manuel; Blouin, Stéphane; Akcan, Emre Rüstü; Findl, Oliver.
Afiliación
  • Stunf Pukl S; Vienna Institute for Research in Ocular Surgery (VIROS), A Karl Landsteiner Institute, Hanusch Hospital, Heinrich-Collin Str. 30, AT-1140, Vienna, Austria.
  • Kronschläger M; Medical Faculty, University of Ljubljana, Ljubljana, Slovenia.
  • Ruiss M; Eye Hospital, University Clinical Centre Ljubljana, Ljubljana, Slovenia.
  • Blouin S; Vienna Institute for Research in Ocular Surgery (VIROS), A Karl Landsteiner Institute, Hanusch Hospital, Heinrich-Collin Str. 30, AT-1140, Vienna, Austria. martin.kronschlaeger@gmx.at.
  • Akcan ER; Vienna Institute for Research in Ocular Surgery (VIROS), A Karl Landsteiner Institute, Hanusch Hospital, Heinrich-Collin Str. 30, AT-1140, Vienna, Austria.
  • Findl O; Ludwig Boltzmann Institute of Osteology at the Hanusch Hospital of OEGK and AUVA Trauma Centre Meidling, Medical Department Hanusch Hospital, Vienna, Austria.
BMC Ophthalmol ; 24(1): 103, 2024 Mar 05.
Article en En | MEDLINE | ID: mdl-38443841
ABSTRACT

PURPOSE:

To measure the dislocation forces in relation to haptic material, flange size and needle used.

SETTING:

Hanusch Hospital, Vienna, Austria.

DESIGN:

Laboratory Investigation. METHODS, MAIN OUTCOME

MEASURES:

30 G (gauge) thin wall and 27 G standard needles were used for a 2 mm tangential scleral tunnel in combination with different PVDF (polyvinylidene fluoride) and PMMA (polymethylmethacrylate haptics). Flanges were created by heating 1 mm of the haptic end, non-forceps assisted in PVDF and forceps assisted in PMMA haptics. The dislocation force was measured in non-preserved cadaver sclera using a tensiometer device.

RESULTS:

PVDF flanges achieved were of a mushroom-like shape and PMMA flanges were of a conic shape. For 30 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 1.58 ± 0.68 N (n = 10) and 0.70 ± 0.14 N (n = 9) (p = 0.003) respectively. For 27 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 0.31 ± 0.35 N (n = 3) and 0.0 N (n = 4), respectively. The flange size correlated with the occurring dislocation force in experiments with 30 G needle tunnels (r = 0.92), when flanges were bigger than 384 micrometres.

CONCLUSIONS:

The highest dislocation forces were found for PVDF haptic flanges and their characteristic mushroom-like shape for 30 G thin wall needle scleral tunnels. Forceps assisted flange creation in PMMA haptics did not compensate the disadvantage of PMMA haptics with their characteristic conic shape flange.