Dislocation force of scleral flange-fixated intraocular lens haptics.
BMC Ophthalmol
; 24(1): 103, 2024 Mar 05.
Article
en En
| MEDLINE
| ID: mdl-38443841
ABSTRACT
PURPOSE:
To measure the dislocation forces in relation to haptic material, flange size and needle used.SETTING:
Hanusch Hospital, Vienna, Austria.DESIGN:
Laboratory Investigation. METHODS, MAIN OUTCOMEMEASURES:
30 G (gauge) thin wall and 27 G standard needles were used for a 2 mm tangential scleral tunnel in combination with different PVDF (polyvinylidene fluoride) and PMMA (polymethylmethacrylate haptics). Flanges were created by heating 1 mm of the haptic end, non-forceps assisted in PVDF and forceps assisted in PMMA haptics. The dislocation force was measured in non-preserved cadaver sclera using a tensiometer device.RESULTS:
PVDF flanges achieved were of a mushroom-like shape and PMMA flanges were of a conic shape. For 30 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 1.58 ± 0.68 N (n = 10) and 0.70 ± 0.14 N (n = 9) (p = 0.003) respectively. For 27 G needle tunnels the dislocation forces for PVDF and PMMA haptic flanges were 0.31 ± 0.35 N (n = 3) and 0.0 N (n = 4), respectively. The flange size correlated with the occurring dislocation force in experiments with 30 G needle tunnels (r = 0.92), when flanges were bigger than 384 micrometres.CONCLUSIONS:
The highest dislocation forces were found for PVDF haptic flanges and their characteristic mushroom-like shape for 30 G thin wall needle scleral tunnels. Forceps assisted flange creation in PMMA haptics did not compensate the disadvantage of PMMA haptics with their characteristic conic shape flange.Palabras clave
Texto completo:
1
Base de datos:
MEDLINE
Asunto principal:
Polivinilos
/
Polímeros de Fluorocarbono
/
Tecnología Háptica
/
Lentes Intraoculares
Idioma:
En
Revista:
BMC Ophthalmol
Asunto de la revista:
OFTALMOLOGIA
Año:
2024
Tipo del documento:
Article