Assessment and validation of glottic motion using cone-beam CT and real-time cine MRI.
Strahlenther Onkol
; 200(5): 418-424, 2024 May.
Article
en En
| MEDLINE
| ID: mdl-38488899
ABSTRACT
PURPOSE:
This study aimed to assess the margin for the planning target volume (PTV) using the Van Herk formula. We then validated the proposed margin by real-time magnetic resonance imaging (MRI).METHODS:
An analysis of cone-beam computed tomography (CBCT) data from early glottic cancer patients was performed to evaluate organ motion. Deformed clinical target volumes (CTV) after rigid registration were acquired using the Velocity program (Varian Medical Systems, Palo Alto, CA, USA). Systematic (Σ) and random errors (σ) were evaluated. The margin for the PTV was defined as 2.5â¯Σâ¯+ 0.7â¯σ according to the Van Herk formula. To validate this margin, we accrued healthy volunteers. Sagittal real-time cine MRI was conducted using the ViewRay system (ViewRay Inc., Oakwood Village, OH, USA). Within the obtained sagittal images, the vocal cord was delineated. The movement of the vocal cord was summed up and considered as the internal target volume (ITV). We then assessed the degree of overlap between the ITV and the PTV (vocal cord plus margins) by calculating the volume overlap ratio, represented as (ITVâ©PTV)/ITV.RESULTS:
CBCTs of 17 early glottic patients were analyzed. Σ and σ were 0.55 and 0.57 for left-right (LR), 0.70 and 0.60 for anterior-posterior (AP), and 1.84 and 1.04 for superior-inferior (SI), respectively. The calculated margin was 1.8â¯mm (LR), 2.2â¯mm (AP), and 5.3â¯mm (SI). Four healthy volunteers participated for validation. A margin of 3â¯mm (AP) and 5â¯mm (SI) was applied to the vocal cord as the PTV. The average volume overlap ratio between ITV and PTV was 0.92 (range 0.85-0.99) without swallowing and 0.77 (range 0.70-0.88) with swallowing.CONCLUSION:
By evaluating organ motion by using CBCT, the margin was 1.8 (LR), 2.2 (AP), and 5.3â¯mm (SI). The margin acquired using CBCT fitted well in real-time cine MRI. Given that swallowing during radiotherapy can result in a substantial displacement, it is crucial to consider strategies aimed at minimizing swallowing and related motion.Palabras clave
Texto completo:
1
Base de datos:
MEDLINE
Asunto principal:
Neoplasias Laríngeas
/
Imagen por Resonancia Cinemagnética
/
Tomografía Computarizada de Haz Cónico
/
Glotis
Idioma:
En
Revista:
Strahlenther Onkol
Asunto de la revista:
NEOPLASIAS
/
RADIOTERAPIA
Año:
2024
Tipo del documento:
Article