SPFS: SNR peak-based frequency selection method to alleviate resolution degradation in MPI real-time imaging.
Shan, Shihao; Zhang, Chenglong; Cheng, Min; Qi, Yafei; Yu, Dexin; Wildgruber, Moritz; Ma, Xiaopeng.
Afiliación
  • Shan S; School of Control Science and Engineering, Shandong University, Jinan, Shandong 250061, People's Republic of China.
  • Zhang C; School of Control Science and Engineering, Shandong University, Jinan, Shandong 250061, People's Republic of China.
  • Cheng M; Xintai hospital of traditional Chinese medicine, Tai'an, Shandong, People's Republic of China.
  • Qi Y; Department of Radiology, Qilu Hospital, Cheeloo College of Medicine, Shandong University, Jinan, Shandong 250012, People's Republic of China.
  • Yu D; Department of Radiology, Qilu Hospital, Cheeloo College of Medicine, Shandong University, Jinan, Shandong 250012, People's Republic of China.
  • Wildgruber M; Department of Radiology, University Hospital, LMU Munich, Munich D-81337, Germany.
  • Ma X; School of Control Science and Engineering, Shandong University, Jinan, Shandong 250061, People's Republic of China.
Phys Med Biol ; 69(11)2024 May 21.
Article en En | MEDLINE | ID: mdl-38593815
ABSTRACT
Objective. The primary objective of this study is to address the reconstruction time challenge in magnetic particle imaging (MPI) by introducing a novel approach named SNR-peak-based frequency selection (SPFS). The focus is on improving spatial resolution without compromising reconstruction speed, thereby enhancing the clinical potential of MPI for real-time imaging.Approach. To overcome the trade-off between reconstruction time and spatial resolution in MPI, the researchers propose SPFS as an innovative frequency selection method. Unlike conventional SNR-based selection, SPFS prioritizes frequencies with signal-to-noise ratio (SNR) peaks that capture crucial system matrix information. This adaptability to varying quantities of selected frequencies enhances versatility in the reconstruction process. The study compares the spatial resolution of MPI reconstruction using both SNR-based and SPFS frequency selection methods, utilizing simulated and real device data.Main results.The research findings demonstrate that the SPFS approach substantially improves image resolution in MPI, especially when dealing with a limited number of frequency components. By focusing on SNR peaks associated with critical system matrix information, SPFS mitigates the spatial resolution degradation observed in conventional SNR-based selection methods. The study validates the effectiveness of SPFS through the assessment of MPI reconstruction spatial resolution using both simulated and real device data, highlighting its potential to address a critical limitation in the field.Significance.The introduction of SPFS represents a significant breakthrough in MPI technology. The method not only accelerates reconstruction time but also enhances spatial resolution, thus expanding the clinical potential of MPI for various applications. The improved real-time imaging capabilities of MPI, facilitated by SPFS, hold promise for advancements in drug delivery, plaque assessment, tumor treatment, cerebral perfusion evaluation, immunotherapy guidance, andin vivocell tracking.
Asunto(s)
Palabras clave

Texto completo: 1 Base de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Relación Señal-Ruido Idioma: En Revista: Phys Med Biol Año: 2024 Tipo del documento: Article