Head size in Delta ceramic-on-ceramic total hip arthroplasty: a comparative registry study.
Castagnini, Francesco; Bordini, Barbara; Cosentino, Monica; Diquattro, Emanuele; Gorgone, Mara; Traina, Francesco.
Afiliación
  • Castagnini F; Ortopedia-Traumatologia e Chirurgia Protesica e dei Reimpianti d'anca e di Ginocchio, IRCCS Istituto Ortopedico Rizzoli, Via Pupilli 1, 40136, Bologna, Italy. francescocastagnini@hotmail.it.
  • Bordini B; Laboratorio di Tecnologia Medica, IRCCS Istituto Ortopedico Rizzoli, Via di Barbiano 1/10, 40136, Bologna, Italy.
  • Cosentino M; Laboratorio di Tecnologia Medica, IRCCS Istituto Ortopedico Rizzoli, Via di Barbiano 1/10, 40136, Bologna, Italy.
  • Diquattro E; Ortopedia-Traumatologia e Chirurgia Protesica e dei Reimpianti d'anca e di Ginocchio, IRCCS Istituto Ortopedico Rizzoli, Via Pupilli 1, 40136, Bologna, Italy.
  • Gorgone M; Ortopedia-Traumatologia e Chirurgia Protesica e dei Reimpianti d'anca e di Ginocchio, IRCCS Istituto Ortopedico Rizzoli, Via Pupilli 1, 40136, Bologna, Italy.
  • Traina F; Head of Ortopedia-Traumatologia e, Chirurgia Protesica e dei Reimpianti d'anca e di Ginocchio, IRCCS Istituto Ortopedico Rizzoli, Via Pupilli 1, 40136, Bologna, Italy.
Article en En | MEDLINE | ID: mdl-39225803
ABSTRACT

INTRODUCTION:

Optimal bearing size in total hip arthroplasty (THA) has not been clearly ascertained, but large diameter ceramic balls may improve function and implant stability, with minimal concerns for wear. Delta ceramic-on-ceramic (COC) THA were stratified based on head diameter (32, 36 and 40 mm), aiming to assess 1) survival rates, reasons for revision and hazard ratios for failure; 2) survival rates and hazard ratios for aseptic loosening, dislocation/primary instability, liner failure. MATERIALS AND

METHODS:

A regional arthroplasty registry was enquired about Delta COC THA performed for primary osteoarthritis, stratified by head size. 13,161 primary cementless THAs were included in the study, with a mean follow-up of 5.7 years (range 0-17) 3980 (30.2%) 32 mm balls, 8327 (63.3%) 36 mm heads, 854 (6.5%) 40 mm heads.

RESULTS:

The three cohorts achieved similar survival rates (p = 0.99) and adjusted hazard ratios (p > 0.05). No revisions for head breakage or metallosis occurred. When revision for cup aseptic loosening was the endpoint, the survival rates of three cohorts (p = 0.08) and the adjusted hazard ratios (p > 0.05) were similar. The three cohorts achieved comparable survival rates when revision for dislocation/primary instability was the endpoint (p = 0.08). When the endpoint was revision for liner breakage, 32 mm cohort showed higher rates of revision (p = 0.01). No liner failure was detected in the 40 mm cohort.

CONCLUSIONS:

32 mm, 36 mm and 40 mm provided similar 10-year survival rates in Delta COC THA. 40 mm heads emerged as safe options but did not lower the revisions for dislocations/primary instability. LEVEL OF EVIDENCE IV (therapeutic study).