RESUMEN
PURPOSE: To investigate the different sized measurement files on the accuracy of electronic apex locator (EAL) in vitro. METHODS: Alginate and extracted single-Ccanalled tooth were used to mimic the situation in vivo. Forty-one teeth were used. EAL Raypex 5 was applied to determine the length of the canals by 10#-40# K files. Experimental measurements and the distances (IF value) between experimental and ideal actual measurements under the same measurement environment were recorded. Paired sample F test was applied to analyze the results by using SPSS 10.0 software package. RESULTS: The results showed that at the allowance of + or - 0.5mm, the accuracy ratios were 85.4%(10#),75.6%(15#),85.4%(20#),90.2%(25#),73.2%(30#),68.3%(35#) and 63.4%(40#), respectively. While at the allowance of +0.5--1.0mm, the accuracy ratios were 87.8%(10#),87.8%(15#),90.2%(20#),97.6%(25#),97.6%(30#),97.6%(35#) and 87.8%(40#), respectively. Statistical analysis showed that there was no significant difference in accuracy between each group(P>0.05). CONCLUSION: The results suggested that it be advisable to improve the accuracy of the EAL by measuring the working length one more time with 25# K file during the preparation under this experimental environment.