1.
Proc Inst Mech Eng H ; 220(8): 929-38, 2006 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-17236525

RESUMEN

This paper reports on the performance evaluation of a dental handpiece in simulation of clinical finishing using a novel two-degrees-of-freedom (2DOF) in vitro apparatus. The instrumented apparatus consisted of a two-dimensional computer-controlled coordinate worktable carrying a dental handpiece, a piezoelectric force dynamometer, and a high-speed data acquisition and signal conditioning system for simulating the clinical operations and monitoring the dental finishing processes. The performance of the dental handpiece was experimentally evaluated with respect to rotational speed, torque, and specific finishing energy under the applied clinical finishing conditions. The results show that the rotational speeds of the dental handpiece decreased by increasing either the depth of cut or the feed rate at a constant clinically applied air pressure and water flowrate. They also decreased when increasing both the tangential and normal finishing forces. The specific finishing energy decreased with an increase in either depth of cut or feed rate, while the finishing torque increased as either the depth of cut or the feed rate was increased. Implications of these results were to provide guidance for proper applications of dental handpieces in clinical practice.


Asunto(s)
Equipo Dental de Alta Velocidad , Pulido Dental/instrumentación , Análisis de Falla de Equipo/instrumentación , Análisis de Falla de Equipo/métodos , Diseño de Equipo , Rotación , Torque
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA