RESUMO

We report the safe and effective use of an ultrasonically activated scalpel as an alternative to unipolar electrocautery during surgical procedures in 4 patients with pacemakers. This scalpel provided adequate hemostasis without the adverse consequences of electromagnetic interference. As the scalpel was able to easily cut through the silicone connector sleeve of one generator, it is not recommended for routine generator replacement because it may damage existing pacing leads. However, its use during nonpacemaker operations in patients with pacemakers should be considered.


Assuntos
Eletrocoagulação , Marca-Passo Artificial , Terapia por Ultrassom/instrumentação , Adulto , Bacteriemia/microbiologia , Infecções Bacterianas/cirurgia , Fístula Cutânea/cirurgia , Eletrocoagulação/instrumentação , Fenômenos Eletromagnéticos , Feminino , Hematoma/cirurgia , Hemostasia Cirúrgica/instrumentação , Humanos , Masculino , Marca-Passo Artificial/efeitos adversos , Segurança , Silicones , Infecções Estafilocócicas , Infecções Cutâneas Estafilocócicas/cirurgia , Procedimentos Cirúrgicos Operatórios