Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros

Base de dados
Ano de publicação
Tipo de documento
Assunto da revista
País de afiliação
Intervalo de ano de publicação
1.
J Shoulder Elbow Surg ; 25(4): 572-80, 2016 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-26577127

RESUMO

BACKGROUND: Numerous studies have documented the concern for progressive radiolucent lines, signifying debonding and subsequent aseptic loosening of the glenoid component. In this study, we compared 3 cementation methods to secure a central peg in 15 cadaveric glenoids. METHODS: Cement application techniques consisted of (1) compression of multiple applications of cement using manual pressure over gauze with an Adson clamp, (2) compression of multiple applications of cement using a pressurizer device, and (3) no compression of a single application of cement. Each glenoid was then imaged with high-resolution micro-computed tomography and further processed by creating 3-dimensional computerized models of implant, bone, and cement geometry. Cement morphology characteristics were then analyzed in each of the models. RESULTS: There were no significant differences detected between the 2 types of compression techniques; however, there was a significant difference between compression methods and use of no compression at all. All morphologic characteristics of a larger cement mantle were significantly correlated with greater cortical contact. CONCLUSIONS: We demonstrate that compression techniques create a larger cement mantle. Increased size of the cement mantle is associated with increased contact with cortical bone at the glenoid vault. This method for characterizing the cement mantle by micro-computed tomography scanning techniques and 3-dimensional analysis may also be useful in future finite element analysis studies.


Assuntos
Artroplastia de Substituição/métodos , Cimentação/métodos , Escápula/diagnóstico por imagem , Articulação do Ombro/diagnóstico por imagem , Articulação do Ombro/cirurgia , Microtomografia por Raio-X , Cimentos Ósseos , Cadáver , Simulação por Computador , Análise de Elementos Finitos , Humanos , Prótese Articular , Pressão , Falha de Prótese , Escápula/cirurgia
2.
J Orthop Res ; 33(11): 1671-9, 2015 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-25929691

RESUMO

Aseptic loosening of cemented joint replacements is a complex biological and mechanical process, and remains a clinical concern especially in patients with poor bone quality. Utilizing high resolution finite element analysis of a series of implanted cadaver glenoids, the objective of this study was to quantify relationships between construct morphology and resulting mechanical stresses in cement and trabeculae. Eight glenoid cadavers were implanted with a cemented central peg implant. Specimens were imaged by micro-CT, and subject-specific finite element models were developed. Bone volume fraction, glenoid width, implant-cortex distance, cement volume, cement-cortex contact, and cement-bone interface area were measured. Axial loading was applied to the implant of each model and stress distributions were characterized. Correlation analysis was completed across all specimens for pairs of morphological and mechanical variables. The amount of trabecular bone with high stress was strongly negatively correlated with both cement volume and contact between the cement and cortex (r = -0.85 and -0.84, p < 0.05). Bone with high stress was also correlated with both glenoid width and implant-cortex distance. Contact between the cement and underlying cortex may dramatically reduce trabecular bone stresses surrounding the cement, and this contact depends on bone shape, cement amount, and implant positioning.


Assuntos
Cimentos Ósseos , Osso e Ossos/diagnóstico por imagem , Falha de Prótese , Estresse Mecânico , Idoso , Análise de Elementos Finitos , Humanos , Microtomografia por Raio-X
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA