J Endovasc Ther ; 21(2): 230-9, 2014 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-24754282

RESUMO

PURPOSE: To investigate the corrosion resistance properties of 5 commercially available nitinol stents used to treat peripheral artery disease and compare their surface quality, elemental composition, and geometrical design. METHODS: Samples of 5 different designs of nitinol peripheral stents [LifeStent (n=4), Philon (n=6), Epic (n=6), S.M.A.R.T. Control (n=7), and Complete SE (n=7)] were examined using stereomicroscopy, environmental scanning electron microscopy, and x-ray photoelectron spectroscopy. Corrosion resistance testing was performed in accordance with ASTM International Standard F2129-08. RESULTS: Thirteen (43%) of 30 stents corroded during this experiment. Stent fracture was observed in 12 (92%) of these corroded stents. Mean breakdown potentials ranged from 517 to 835 mV (vs. Ag/AgCl) for the Philon, Complete SE, S.M.A.R.T. Control, Epic, and LifeStent models from lowest to highest. A statistically significant difference in breakdown potential was observed between the LifeStent vs. Philon stents (835 vs. 517 mV, p=0.01) and Epic vs. Philon stents (833 vs. 517 mV, p=0.03). Stents with lower breakdown potential and relative breakdown potentials were associated with a higher fracture frequency (Spearman correlation coefficient -0.44, p=0.015 and -0.869, p<0.01, respectively). CONCLUSION: In this in vitro study, corrosion led independently to stent fracture. There is a significant association between lower mean breakdown/relative breakdown potentials and stent fracture.


Assuntos
Ligas , Procedimentos Endovasculares/instrumentação , Doença Arterial Periférica/terapia , Desenho de Prótese , Falha de Prótese , Stents , Corrosão , Humanos , Teste de Materiais , Microscopia Eletrônica de Varredura , Espectroscopia Fotoeletrônica , Propriedades de Superfície
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA