RESUMO

PURPOSE: To evaluate the safety and efficacy of the Igaki-Tamai biodegradable scaffold after drug-eluting balloon (DEB) angioplasty in patients with occlusive superficial femoral artery (SFA) disease. METHODS: A prospective, single-center, nonrandomized study enrolled 20 patients (mean age 66.7±11.6 years; 14 men) with symptomatic de novo SFA lesions undergoing angioplasty with the In.Pact Admiral paclitaxel-coated balloon and subsequent implantation of the Igaki-Tamai bioresorbable scaffold. All patients were claudicants. The average diameter stenosis was 89.7%, and the mean length was 43.6 mm. Clinical examinations with duplex sonography were performed after 1, 6, 9, and 12 months. The main study outcomes were technical success, restenosis, target lesion revascularization (TLR), ankle-brachial index (ABI) improvement, and changes in quality of life evaluated with the walking impairment questionnaire. Safety was assessed by monitoring the occurrence of adverse events. RESULTS: Angioplasty with a paclitaxel-coated balloon was performed in all patients, resulting in an average diameter stenosis of 24%. Subsequent implantation of the Igaki-Tamai scaffold reduced the average diameter stenosis to 3.5%. In the first 6 months, 2 cases of restenosis were reported, with no TLRs within that period. However, by the 12-month follow-up in 19 patients, 11 patients had lost in-stent patency. Among these patients, 8 had TLRs, which were the only adverse events recorded that were referable to the procedure. Quality-of-life assessments showed improvement in the majority of patients. CONCLUSION: The GAIA-DEB study shows that DEB treatment of the femoral artery prior to the implantation of the biodegradable Igaki-Tamai scaffold is safe. However, the antiproliferative actions of paclitaxel in the vessel wall were not effective in preventing restenosis. In-stent restenosis occurred predominantly after 6 months.


Assuntos
Implantes Absorvíveis , Angioplastia com Balão/instrumentação , Fármacos Cardiovasculares/administração & dosagem , Materiais Revestidos Biocompatíveis , Artéria Femoral , Paclitaxel/administração & dosagem , Doença Arterial Periférica/terapia , Dispositivos de Acesso Vascular , Idoso , Angioplastia com Balão/efeitos adversos , Índice Tornozelo-Braço , Constrição Patológica , Feminino , Artéria Femoral/diagnóstico por imagem , Artéria Femoral/fisiopatologia , Alemanha , Humanos , Masculino , Pessoa de Meia-Idade , Doença Arterial Periférica/diagnóstico , Doença Arterial Periférica/fisiopatologia , Projetos Piloto , Estudos Prospectivos , Recidiva , Retratamento , Inquéritos e Questionários , Fatores de Tempo , Resultado do Tratamento , Ultrassonografia Doppler Dupla , Grau de Desobstrução Vascular
2.
J Endovasc Ther ; 18(6): 745-52, 2011 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-22149221

RESUMO

PURPOSE: To examine the efficacy and integrity of a novel interwoven self-expanding nitinol stent system for the treatment of complex femoropopliteal lesions in a "real world" medical practice. METHODS: This retrospective analysis included 107 consecutive patients (77 men; mean age 68.9 years) with atherosclerotic femoropopliteal lesions (occlusions in 31%) who underwent implantation of 137 SUPERA stents. The patients were followed for up to 24 months by Doppler ultrasound examinations, radiography of the stent, and assessments of Rutherford-Becker class and ankle-brachial index (ABI). RESULTS: The mean implanted stent length was 111±50 mm (range 40-270). Procedure success (residual stenosis <30%) was achieved in 99% of procedures. The 6-, 12-, and 24-month cumulative primary patency rates (± standard error) were 93.1%±2.5%, 84.7%±3.6%, and 76.1%±4.5%, respectively, and the secondary patency rates were 99.0%±0.1%, 94.8%±0.2% and 91.9%±0.3%, respectively. Between baseline and 24 months, mean ABI increased from 0.68±0.14 to 0.87±0.10 and the mean Rutherford-Becker class decreased from 3.3±0.7 to 2.0±1.0 (p<0.0001 for both). Radiographs performed in 91 patients at a mean of 16.8±7.0 months found no stent fractures. CONCLUSION: Over a 2-year surveillance period, excellent durability without stent fractures was documented after implantation of the SUPERA stent in complex femoropopliteal lesions. In addition, significant improvements were observed in symptom classification and hemodynamics.


Assuntos
Artéria Femoral/patologia , Doença Arterial Periférica/terapia , Artéria Poplítea/patologia , Stents , Adulto , Idoso , Idoso de 80 Anos ou mais , Ligas , Feminino , Seguimentos , Humanos , Masculino , Pessoa de Meia-Idade , Desenho de Prótese , Estudos Retrospectivos , Resultado do Tratamento
3.
J Am Coll Cardiol ; 45(2): 312-5, 2005 Jan 18.
Artigo em Inglês | MEDLINE | ID: mdl-15653033

RESUMO

OBJECTIVES: The aim of this study was to investigate the occurrence and the clinical impact of stent fractures after femoropopliteal stenting. BACKGROUND: The development of femoral stent fractures has recently been described; however, there are no data about the frequency and the clinical relevance. METHODS: A systematic X-ray screening for stent fractures was performed in 93 patients. In total, 121 legs treated by implantation of self-expanding nitinol stents were investigated after a mean follow-up time of 10.7 months. The mean length of the stented segment was 15.7 cm. RESULTS: Overall, stent fractures were detected in 45 of 121 treated legs (37.2%). In a stent-based analysis, 64 of 261 stents (24.5%) showed fractures, which were classified as minor (single strut fracture) in 31 cases (48.4%), moderate (fracture of >1 strut) in 17 cases (26.6%), and severe (complete separation of stent segments) in 16 cases (25.0%). Fracture rates were 13.2% for stented length < or =8 cm, 42.4% for stented length >8 to 16 cm, and 52.0% for stented length >16 cm. In 21 cases (32.8%) there was a restenosis of >50% diameter reduction at the site of stent fracture. In 22 cases (34.4%) with stent fracture there was a total stent reocclusion. According to Kaplan-Meier estimates, the primary patency rate at 12 months was significantly lower for patients with stent fractures (41.1% vs. 84.3%, p < 0.0001). CONCLUSIONS: There is a considerable risk of stent fractures after long segment femoral artery stenting, which is associated with a higher in-stent restenosis and reocclusion rate.


Assuntos
Arteriopatias Oclusivas/cirurgia , Artéria Femoral/diagnóstico por imagem , Oclusão de Enxerto Vascular/etiologia , Artéria Poplítea/diagnóstico por imagem , Stents/efeitos adversos , Ligas , Arteriopatias Oclusivas/diagnóstico por imagem , Falha de Equipamento , Artéria Femoral/cirurgia , Seguimentos , Oclusão de Enxerto Vascular/diagnóstico por imagem , Humanos , Artéria Poplítea/cirurgia , Radiografia , Índice de Gravidade de Doença , Grau de Desobstrução Vascular
4.
JACC Cardiovasc Interv ; 9(7): 715-24, 2016 Apr 11.
Artigo em Inglês | MEDLINE | ID: mdl-27056311

RESUMO

OBJECTIVES: The authors sought to investigate the efficacy of a drug-coated balloon (DCB) for treatment of complex femoropopliteal lesions. BACKGROUND: Superiority of DCBs compared with uncoated balloon angioplasty for femoropopliteal interventions has been demonstrated in randomized trials for short lesions. Their performance in complex lesions with higher restenosis rates is unclear. METHODS: Patency, target lesion revascularization (TLR) rate, clinical improvement, and safety endpoints of femoropopliteal lesions in 288 limbs (n = 260) treated with the In.Pact Pacific or Admiral DCB (Medtronic, Minneapolis, Minnesota) were retrospectively analyzed for up to 2 years of follow-up. Predictors of restenosis were identified by logistic regression. RESULTS: Lesions were de novo in 51.7%, restenosis in 11.1%, and in-stent restenosis in 37.2%. Mean lesion length was 24.0 ± 10.2 cm, and 65.3% were occluded. Stent implantation was performed in 23.3%. Kaplan Meier estimates of primary patency were 79.2% and 53.7% for all lesions at 1 and 2 years, respectively, whereas freedom from TLR was 85.4% and 68.6%. Primary patency for in-stent restenosis treatment was 76.6% and 48.6%, and freedom from TLR was 83.0% and 58.7% at 1 and 2 years, respectively. Rutherford category improved from a median 3.3 to 1.2 at 1 year, and to 1.1 at 2 years. Major amputation rate was 2.1% at 2 years. No adverse events were thought to be attributable to the coating of the balloon. CONCLUSIONS: These results suggest that DCB are safe and effective in delaying rather than preventing restenosis in long, complex lesions and restenosis of the femoropopliteal tract. Further studies are recommended to confirm these results.


Assuntos
Angioplastia com Balão/instrumentação , Fármacos Cardiovasculares/administração & dosagem , Materiais Revestidos Biocompatíveis , Artéria Femoral/efeitos dos fármacos , Doença Arterial Periférica/terapia , Artéria Poplítea/efeitos dos fármacos , Dispositivos de Acesso Vascular , Idoso , Angioplastia com Balão/efeitos adversos , Fármacos Cardiovasculares/efeitos adversos , Distribuição de Qui-Quadrado , Desenho de Equipamento , Feminino , Artéria Femoral/diagnóstico por imagem , Artéria Femoral/fisiopatologia , Humanos , Estimativa de Kaplan-Meier , Modelos Logísticos , Masculino , Pessoa de Meia-Idade , Razão de Chances , Doença Arterial Periférica/diagnóstico por imagem , Doença Arterial Periférica/fisiopatologia , Artéria Poplítea/diagnóstico por imagem , Artéria Poplítea/fisiopatologia , Recidiva , Sistema de Registros , Estudos Retrospectivos , Fatores de Risco , Stents , Fatores de Tempo , Resultado do Tratamento , Grau de Desobstrução Vascular
5.
JACC Cardiovasc Interv ; 6(1): 65-71, 2013 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-23347863

RESUMO

OBJECTIVES: We examined the efficacy and durability of a new interwoven self-expanding nitinol stent system in the treatment of complex popliteal artery lesions in unselected patients. BACKGROUND: The optimal endovascular treatment strategy for atherosclerotic popliteal artery disease is not known. METHODS: We retrospectively analyzed the data gathered in 101 consecutive patients presenting with atherosclerotic, popliteal arterial disease, who underwent implantation of 125 stents. The patients were followed for 12 months by Doppler ultrasound examinations, stent roentgenograms, and estimation of Rutherford-Becker class (RBC) and ankle-brachial index (ABI). RESULTS: The mean age of the patients was 73.1 years, and 52.5% were men. Total occlusions were present in 48 patients (47.5%). The mean stent length was 84.3 ± 45.1 mm (range 40 to 240 mm). A <30% residual stenosis was achieved in 98.0% of procedures. The 6- and 12-month primary patency rates were 94.6 ± 2.3% and 87.7 ± 3.7%, respectively, and the secondary patency rates 97.9 ± 1.5% and 96.5 ± 2.0%, respectively. Between baseline and 12 months of follow-up, mean ABI increased from 0.58 ± 0.15 to 0.97 ± 0.18, and mean RBC decreased from 3.1 ± 0.9 to 1.4 ± 0.8 (p < 0.001 for both comparisons). Radiographs performed on 51 patients, at a mean of 15.2 months, confirmed the absence of stent fractures in 100% of examinations. CONCLUSIONS: Over a 12-month observation period, the patency rate and durability of SUPERA stents implanted for severe popliteal artery disease were high.


Assuntos
Ligas , Angioplastia/instrumentação , Doença Arterial Periférica/terapia , Artéria Poplítea , Stents , Idoso , Idoso de 80 Anos ou mais , Angioplastia/efeitos adversos , Angioplastia/mortalidade , Índice Tornozelo-Braço , Constrição Patológica , Feminino , Alemanha , Humanos , Estimativa de Kaplan-Meier , Masculino , Pessoa de Meia-Idade , Doença Arterial Periférica/diagnóstico , Doença Arterial Periférica/mortalidade , Doença Arterial Periférica/fisiopatologia , Artéria Poplítea/diagnóstico por imagem , Artéria Poplítea/fisiopatologia , Valor Preditivo dos Testes , Desenho de Prótese , Radiografia , Sistema de Registros , Estudos Retrospectivos , Índice de Gravidade de Doença , Fatores de Tempo , Resultado do Tratamento , Ultrassonografia Doppler , Grau de Desobstrução Vascular
SELEÇÃO DE REFERÊNCIAS