SHAPE ESTIMATION OF BOWTIE FILTERS BASED ON THE LUMINESCENCE FROM POLYETHYLENE TEREPHTHALATE RESIN BY X-RAY IRRADIATION.
Radiat Prot Dosimetry
; 185(4): 432-439, 2019 Dec 31.
Article
em En
| MEDLINE
| ID: mdl-30916354
ABSTRACT
In this study, we devised a novel method estimating the bowtie filter shapes by imaging luminescence from a polyethylene terephthalate (PET) resin with X-ray irradiation in a computed tomography (CT) scanner. The luminescence distribution of the PET resin corresponding to the thickness of bowtie filter was imaged using a charge-coupled device camera. On the assumption that the material of bowtie filter is aluminium (Al), the shape of bowtie filters was estimated from the correlation between Al attenuation curves and the angular-dependent luminance attenuation profiles according to the thickness of bowtie filters. Dose simulations based on the estimated bowtie filter shapes were performed using head and body PMMA phantoms with 16 and 32 cm in diameter. The simulated values of head and body weighted CT dose index (CTDIw) based on bowtie filter shape by the luminescence imaging method agreed within ~9% with the measured values by a dosemeter.