Influence of artificial aging on mechanical properties of commercially and non-commercially available zirconia dental implants.
Monzavi, Mona; Zhang, Fei; Meille, Sylvain; Douillard, Thierry; Adrien, Jérôme; Noumbissi, Sammy; Nowzari, Hessam; Chevalier, Jérôme.
Afiliação
  • Monzavi M; Private Practice, 120 S Spalding Drive, Beverly Hills, CA, 90212, USA; Université de Lyon, INSA-Lyon, UMR CNRS 5510 MATEIS, 7 Avenue Jean Capelle, 69621, Villeurbanne Cedex, France. Electronic address: monzavi.mona@gmail.com.
  • Zhang F; Université de Lyon, INSA-Lyon, UMR CNRS 5510 MATEIS, 7 Avenue Jean Capelle, 69621, Villeurbanne Cedex, France.
  • Meille S; Université de Lyon, INSA-Lyon, UMR CNRS 5510 MATEIS, 7 Avenue Jean Capelle, 69621, Villeurbanne Cedex, France.
  • Douillard T; Université de Lyon, INSA-Lyon, UMR CNRS 5510 MATEIS, 7 Avenue Jean Capelle, 69621, Villeurbanne Cedex, France.
  • Adrien J; Université de Lyon, INSA-Lyon, UMR CNRS 5510 MATEIS, 7 Avenue Jean Capelle, 69621, Villeurbanne Cedex, France.
  • Noumbissi S; International Academy of Ceramic Implantology Zirconia Implant Research Group-ZIRG, 801 Wayne Avenue, Suite G200, Silver Spring, MD, 20910, USA.
  • Nowzari H; Private Practice, 120 S Spalding Drive, Beverly Hills, CA, 90212, USA.
  • Chevalier J; Université de Lyon, INSA-Lyon, UMR CNRS 5510 MATEIS, 7 Avenue Jean Capelle, 69621, Villeurbanne Cedex, France.
J Mech Behav Biomed Mater ; 101: 103423, 2020 01.
Article em En | MEDLINE | ID: mdl-31536885
ABSTRACT

OBJECTIVE:

To evaluate the effect of artificial aging on the mechanical resistance and micromechanical properties of commercially and noncommercially available zirconia dental implants.

METHODS:

Scanning electron microscopy (SEM) and X-ray computed tomography (X-CT) were performed on implant systems including Z-systems®, Straumann®, Zibone® and commercially and non-commercially available TAV dental® with varying grain sizes. Accelerated aging was performed at 134 °C and 2-bar pressure for 30 hours. Before and after aging, the mechanical load to failure was investigated and the bending moments were calculated. Nanoindentation responses of the representative Zibone implant before and after aging were performed to evaluate the effects of aging on hardness (H) and Young's modulus (E). A two-sample t-test statistical analysis was used to determine significant differences of bending moments within groups.

RESULTS:

All implants presented with compact and homogenous core structures without porosities. The bending moment was significantly increased after aging for all groups (P ≤ 0.05) except for Z-systems (significant decrease (P = 0.022)) and TAV group 3 (no significant increase (P = 0.181)). The increase in bending moment was less pronounced with increasing grain size in TAV groups (group 1 P = 0.036, group 2 P = 0.05, group 3 P = 0.18). E and H were reduced approximately 32% and 18% respectively following aging within the transformed, microcracked zone of the presentative Zibone implant.

CONCLUSIONS:

Aging led to both increase and decrease of the mechanical properties of the implant systems analyzed. The apparent contrast amongst groups can be explained based on differences in grain sizes and surface features. Aging decreased micromechanical properties of one implant system which warrants further investigation.