Health Phys ; 112(3): 237-245, 2017 03.
Artículo en Inglés | MEDLINE | ID: mdl-28121723

RESUMEN

Optically-stimulated luminescent dosimeters are capable of being interrogated multiple times post-irradiation. Each interrogation removes a fraction of the signal stored within the optically-stimulated luminescent dosimeter. This signal loss must be corrected to avoid systematic errors in estimating the average signal of a series of optically-stimulated luminescent dosimeter interrogations and requires a minimum number of consecutive readings to determine an average signal that is within a desired accuracy of the true signal with a desired statistical confidence. This paper establishes a technical basis for determining the required number of readings for a particular application of these dosimeters when using certain OSL dosimetry systems.


Asunto(s)
Artefactos , Guías como Asunto , Iluminación/normas , Dispositivos Ópticos/normas , Dosimetría Termoluminiscente/instrumentación , Dosimetría Termoluminiscente/normas , Calibración/normas , Diseño de Equipo , Análisis de Falla de Equipo , Iluminación/instrumentación , Dosis de Radiación , Reproducibilidad de los Resultados , Semiconductores , Sensibilidad y Especificidad