1.
Pediatr Radiol ; 48(7): 992-998, 2018 07.
Artículo en Inglés | MEDLINE | ID: mdl-29594440

RESUMEN

BACKGROUND: The effects and potential hazards of brain magnetic resonance imaging (MRI) at 3 T in newborns are debated. OBJECTIVE: Assess the impact of 3-T MRI in newborns on body temperature and physiological parameters. MATERIAL AND METHODS: Forty-nine newborns, born preterm and at term, underwent 3-T brain MRI at term-corrected age. Rectal and skin temperature, oxygen saturation and heart rate were recorded before, during and after the scan. RESULTS: A statistically significant increase in skin temperature of 0.6 °C was observed at the end of the MRI scan (P<0.01). There was no significant changes in rectal temperature, heart rate or oxygen saturation. CONCLUSION: Core temperature, heart rate and oxygen saturation in newborns were not affected by 3-T brain MR scanning.


Asunto(s)
Temperatura Corporal , Encéfalo/diagnóstico por imagen , Imagen por Resonancia Magnética/métodos , Seguridad del Paciente , Femenino , Frecuencia Cardíaca , Humanos , Recién Nacido , Recien Nacido Prematuro , Masculino , Oxígeno/metabolismo , Estudios Prospectivos