Confiabilidad de un monitor automatizado para la medición de la presión arterial / Reliability of an automatic monitor for blood pressure measurement
Cardona-Müller, David; Grover-Páez, Fernando; Guzmán-Saldivar, Víctor; Alanis-Sánchez, Guillermo A; Murguía-Soto, César; Totsuka-Sutto, Sylvia E; Quezada-Fernández, Patricia; Macías-Chumacera, Agustín; Cardona-Muñoz, Ernesto G; Ramos-Becerra, Carlos G.
Afiliación
  • Cardona-Müller, David; Universidad de Guadalajara. Departamento de Fisiología del Centro Universitario de Ciencias de la Salud. Instituto de Terapéutica Experimental y Clínica. Guadalajara. MX
  • Grover-Páez, Fernando; Universidad de Guadalajara. Departamento de Fisiología del Centro Universitario de Ciencias de la Salud. Instituto de Terapéutica Experimental y Clínica. Guadalajara. MX
  • Guzmán-Saldivar, Víctor; Universidad de Guadalajara. Departamento de Fisiología del Centro Universitario de Ciencias de la Salud. Instituto de Terapéutica Experimental y Clínica. Guadalajara. MX
  • Alanis-Sánchez, Guillermo A; Universidad de Guadalajara. Departamento de Fisiología del Centro Universitario de Ciencias de la Salud. Instituto de Terapéutica Experimental y Clínica. Guadalajara. MX
  • Murguía-Soto, César; Universidad de Guadalajara. Departamento de Fisiología del Centro Universitario de Ciencias de la Salud. Instituto de Terapéutica Experimental y Clínica. Guadalajara. MX
  • Totsuka-Sutto, Sylvia E; Universidad de Guadalajara. Departamento de Fisiología del Centro Universitario de Ciencias de la Salud. Instituto de Terapéutica Experimental y Clínica. Guadalajara. MX
  • Quezada-Fernández, Patricia; Universidad de Guadalajara. Departamento de Fisiología del Centro Universitario de Ciencias de la Salud. Instituto de Terapéutica Experimental y Clínica. Guadalajara. MX
  • Macías-Chumacera, Agustín; Universidad de Guadalajara. Departamento de Fisiología del Centro Universitario de Ciencias de la Salud. Instituto de Terapéutica Experimental y Clínica. Guadalajara. MX
  • Cardona-Muñoz, Ernesto G; Universidad de Guadalajara. Departamento de Fisiología del Centro Universitario de Ciencias de la Salud. Instituto de Terapéutica Experimental y Clínica. Guadalajara. MX
  • Ramos-Becerra, Carlos G; Universidad de Guadalajara. Departamento de Fisiología del Centro Universitario de Ciencias de la Salud. Instituto de Terapéutica Experimental y Clínica. Guadalajara. MX
Rev. méd. Chile ; 146(2): 190-195, feb. 2018. tab, graf
Article en Es | LILACS | ID: biblio-961377
Biblioteca responsable: CL1.1
ABSTRACT
Background A correct blood pressure (BP) measurement is essential for the diagnosis and control of high BP.

Aim:

To evaluate the agreement and repeatability of BP measurements with the OMRON HEM-7320-LA device compared to a mercury sphygmomanometer. Material and

Methods:

A cross-sectional study comparing BP measurements made by two randomly selected trained nurses and an automatic oscillometric device. The mercurial sphygmomanometer was connected to the automated device via a "T" type connector and a dual-head stethoscope was used, allowing simultaneous measurements. The results were analyzed with one-factor analysis of variance, Bland-Altman's test, repeatability coefficient (RC), and intra-class correlation coefficient (ICC).

Results:

Forty-nine participants aged 56 ± 19 years were included. Nineteen had hypertension (38%). We did not observe a significant difference in either systolic (SBP) or diastolic blood pressure (DBP) pressure measurements between the observers and the device. The mean difference was −0.09 mmHg (95% confidence intervals (CI)-0.9 to 0.7) for SBP and −0.9 mmHg (95% CI −1.7 to −0.13) for DBP. The RC for SBP (6.2, 5.2 and 5.8 mmHg) and DBP (4.7, 4.2 y 5.2 mmHg) was similar between the observers and the device. The ICC for SBP was 0.990 (95% CI 0.983 to 0.995, p < 0.01) and 0.986 (95% CI 0.977 to 0.991, p < 0.01) for DBP.

Conclusions:

There was a high level of agreement and similar measurement repeatability in the measurements performed by the automatic device and the mercurial sphygmomanometer. No differences in BP measurements were observed.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: LILACS Asunto principal: Determinación de la Presión Sanguínea / Monitores de Presión Sanguínea / Hipertensión Tipo de estudio: Observational_studies / Prevalence_studies / Risk_factors_studies Límite: Female / Humans / Male Idioma: Es Revista: Rev. méd. Chile Asunto de la revista: MEDICINA Año: 2018 Tipo del documento: Article País de afiliación: México

Texto completo: 1 Banco de datos: LILACS Asunto principal: Determinación de la Presión Sanguínea / Monitores de Presión Sanguínea / Hipertensión Tipo de estudio: Observational_studies / Prevalence_studies / Risk_factors_studies Límite: Female / Humans / Male Idioma: Es Revista: Rev. méd. Chile Asunto de la revista: MEDICINA Año: 2018 Tipo del documento: Article País de afiliación: México