A monthly quality assurance procedure for 3D surface imaging.
Wooten, H Omar; Klein, Eric E; Gokhroo, Garima; Santanam, Lakshmi.
Afiliación
  • Wooten HO; Department of Radiation Oncology, University of California, Davis, CA, USA. hasani.wooten@ucdmc.ucdavis.edu
J Appl Clin Med Phys ; 12(1): 3338, 2010 Dec 21.
Article en En | MEDLINE | ID: mdl-21330981
A procedure for periodic quality assurance of a video surface imaging system is introduced. AlignRT is a video camera-based patient localization system that captures and compares images of a patient's topography to a DICOM-formatted external contour, then calculates shifts required to accurately reposition the patient. This technical note describes the tools and methods implemented in our department to verify correct and accurate operation of the AlignRT hardware and software components. The procedure described is performed monthly and complements a daily calibration of the system.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Imagenología Tridimensional Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Planificación de la Radioterapia Asistida por Computador / Imagenología Tridimensional Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos