Simultaneous determination of 143Nd/144Nd and 147Sm/144Nd ratios and Sm-Nd contents from the same filament loaded with purified Sm-Nd aliquot from geological samples by isotope dilution thermal ionization mass spectrometry.
Anal Chem
; 84(14): 6040-7, 2012 Jul 17.
Article
en En
| MEDLINE
| ID: mdl-22746207
Isotope dilution thermal ionization mass spectrometry (ID-TIMS) is the standard technique used to achieve precise (143)Nd/(144)Nd and (147)Sm/(144)Nd isotope ratios and accurate elemental concentrations of Sm-Nd. However, in previous studies, purified Sm and Nd fractions must be individually loaded onto different filaments for their accurate determination using TIMS because of severe isobaric interferences. Thus, the classical ID-TIMS technique is time consuming and laborious. In this study, a new method is proposed, which is able to acquire both ratios of (143)Nd/(144)Nd and (147)Sm/(144)Nd and concentrations of Sm-Nd simultaneously on the same filament arrangement. The measurement time and filament consumption are reduced by 50% with the current method, and therefore, the operation cost of TIMS is significantly reduced. A mixed (152)Sm-(148)Nd spike was employed to achieve accurate results after spike subtraction and isobaric interference corrections. Results obtained from a series of standard rock samples are in good agreement with recommended values, within ±0.003% for the (143)Nd/(144)Nd ratio and ±1% for the (147)Sm/(144)Nd ratio.
Texto completo:
1
Banco de datos:
MEDLINE
Idioma:
En
Revista:
Anal Chem
Año:
2012
Tipo del documento:
Article
País de afiliación:
China