Noise-optimized virtual monoenergetic images and iodine maps for the detection of venous thrombosis in second-generation dual-energy CT (DECT): an ex vivo phantom study.
Bongers, Malte N; Schabel, Christoph; Krauss, Bernhard; Tsiflikas, Ilias; Ketelsen, Dominik; Mangold, Stefanie; Claussen, Claus D; Nikolaou, Konstantin; Thomas, Christoph.
Afiliación
  • Bongers MN; Department of Diagnostic and Interventional Radiology, University Hospital of Tübingen, Hoppe-Seyler-Strasse 3, 72076, Tübingen, Germany.
Eur Radiol ; 25(6): 1655-64, 2015 Jun.
Article en En | MEDLINE | ID: mdl-25528670
ABSTRACT
AIMS AND

OBJECTIVES:

Deep venous thrombosis (DVT) can be difficult to detect using CT due to poor and heterogeneous contrast. Dual-energy CT (DECT) allows iodine contrast optimization using noise-optimized monoenergetic extrapolations (MEIs) and iodine maps (IMs). Our aim was to assess whether MEI and IM could improve the delineation of thrombotic material within iodine-enhanced blood compared to single-energy CT (SECT). MATERIALS AND

METHODS:

Six vessel phantoms, including human thrombus and contrast media-enhanced blood and one phantom without contrast, were placed in an attenuation phantom and scanned with DECT 100/140 kV and SECT 120 kV. IM, virtual non-contrast images (VNC), mixed images, and MEI were calculated. Attenuation of thrombi and blood were measured. Contrast and contrast-to-noise-ratios (CNRs) were calculated and compared among IM, VNC, mixed images, MEI, and SECT using paired t tests.

RESULTS:

MEI40keV and IM showed significantly higher contrast and CNR than SE120kV from high to intermediate iodine concentrations (contrastpMEI40keV < 0.002,pIM < 0.005;CNRpMEI40keV < 0.002,pIM < 0.004). At low iodine concentrations, MEI190keV and VNC images showed significantly higher contrast and CNR than SE120kV with inverted contrasts (contrastpMEI190keV < 0.008,pVNC < 0.002;CNRpMEI190keV < 0.003,pVNC < 0.002).

CONCLUSIONS:

Noise-optimized MEI and IM provide significantly higher contrast and CNR in the delineation of thrombosis compared to SECT, which may facilitate the detection of DVT in difficult cases. KEY POINTS • Poor contrast makes it difficult to detect thrombosis in CT. • Dual-energy-CT allows contrast optimization using monoenergetic extrapolations (MEI) and iodine maps (IM). • Noise-optimized-MEI and IM are significantly superior to single-energy-CT in delineation of thrombosis. • Noise-optimized-MEI and IM may facilitate the detection of deep vein thrombosis.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Intensificación de Imagen Radiográfica / Tomografía Computarizada por Rayos X / Fantasmas de Imagen / Trombosis de la Vena / Yodo Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: Eur Radiol Asunto de la revista: RADIOLOGIA Año: 2015 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Intensificación de Imagen Radiográfica / Tomografía Computarizada por Rayos X / Fantasmas de Imagen / Trombosis de la Vena / Yodo Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: Eur Radiol Asunto de la revista: RADIOLOGIA Año: 2015 Tipo del documento: Article País de afiliación: Alemania