Tremor Frequency Assessment by iPhone® Applications: Correlation with EMG Analysis.
J Parkinsons Dis
; 6(4): 717-721, 2016 10 19.
Article
en En
| MEDLINE
| ID: mdl-27662333
ABSTRACT
Tremor frequency analysis is usually performed by EMG studies but accelerometers are progressively being more used. The iPhone® contains an accelerometer and many applications claim to be capable of measuring tremor frequency. We tested three applications in twenty-two patients with a diagnosis of PD, ET and Holmes' tremor. EMG needle assessment as well as accelerometry was performed at the same time. There was very strong correlation (Pearson >0.8, pâ<â0.001) between the three applications, the EMG needle and the accelerometry. Our data suggests the apps LiftPulse®, iSeismometer® and Studymytremor® are a reliable alternative to the EMG for tremor frequency assessment.