Your browser doesn't support javascript.
loading
Comparing the McGrath Mac Video Laryngoscope and Direct Laryngoscopy for Prehospital Emergency Intubation in Air Rescue Patients: A Multicenter, Randomized, Controlled Trial.
Kreutziger, Janett; Hornung, Sonja; Harrer, Clemens; Urschl, Wilhelm; Doppler, Reinhard; Voelckel, Wolfgang G; Trimmel, Helmut.
Afiliación
  • Kreutziger J; Department of General and Surgical Critical Care Medicine, Medical University of Innsbruck, Innsbruck, Austria.
  • Hornung S; Department of Anesthesiology, Emergency and Critical Care Medicine, Wiener Neustadt General Hospital, Wiener Neustadt, Austria.
  • Harrer C; Karl Landsteiner Institute of Emergency Medicine, Wiener Neustadt General Hospital, Wiener Neustadt, Austria.
  • Urschl W; OEAMTC Air Rescue, Vienna, Austria.
  • Doppler R; OEAMTC Air Rescue, Vienna, Austria.
  • Voelckel WG; OEAMTC Air Rescue, Vienna, Austria.
  • Trimmel H; OEAMTC Air Rescue, Vienna, Austria.
Crit Care Med ; 47(10): 1362-1370, 2019 10.
Article en En | MEDLINE | ID: mdl-31389835
ABSTRACT

OBJECTIVES:

Tracheal intubation in prehospital emergency care is challenging. The McGrath Mac Video Laryngoscope (Medtronic, Minneapolis, MN) has been proven to be a reliable alternative for in-hospital airway management. This trial compared the McGrath Mac Video Laryngoscope and direct laryngoscopy for the prehospital setting.

DESIGN:

Multicenter, prospective, randomized, controlled equivalence trial.

SETTING:

Oesterreichischer Automobil- und Touring Club (OEAMTC) Helicopter Emergency Medical Service in Austria, 18-month study period. PATIENTS Five-hundred fourteen adult emergency patients (≥ 18 yr old).

INTERVENTIONS:

Helicopter Emergency Medical Service physicians followed the institutional algorithm, comprising a maximum of two tracheal intubation attempts with each device, followed by supraglottic, then surgical airway access in case of tracheal intubation failure. No restrictions were given for tracheal intubation indication. MEASUREMENTS MAIN

RESULTS:

The Primary outcome was the rate of successful tracheal intubation; equivalence range was ± 6.5% of success rates. Secondary outcomes were the number of attempts to successful tracheal intubation, time to glottis passage and first end-tidal CO2 measurement, degree of glottis visualization, and number of problems. The success rate for the two devices was equivalent direct laryngoscopy 98.5% (254/258), McGrath Mac Video Laryngoscope 98.1% (251/256) (difference, 0.4%; 99% CI, -2.58 to 3.39). There was no statistically significant difference with regard to tracheal intubation times, number of attempts or difficulty. The view to the glottis was significantly better, but the number of technical problems was increased with the McGrath Mac Video Laryngoscope. After a failed first tracheal intubation attempt, immediate switching of the device was significantly more successful than after the second attempt (90.5% vs 57.1%; p = 0.0003), regardless of the method.

CONCLUSIONS:

Both devices are equivalently well suited for use in prehospital emergency tracheal intubation of adult patients. Switching the device following a failed first tracheal intubation attempt was more successful than a second attempt with the same device.
Asunto(s)

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Servicios Médicos de Urgencia / Intubación Intratraqueal / Laringoscopía Tipo de estudio: Clinical_trials / Observational_studies Límite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Crit Care Med Año: 2019 Tipo del documento: Article País de afiliación: Austria

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Servicios Médicos de Urgencia / Intubación Intratraqueal / Laringoscopía Tipo de estudio: Clinical_trials / Observational_studies Límite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Crit Care Med Año: 2019 Tipo del documento: Article País de afiliación: Austria