Your browser doesn't support javascript.
loading
Determining the longitudinal accuracy and reproducibility of T1 and T2 in a 3T MRI scanner.
Carr, Madeline E; Keenan, Kathryn E; Rai, Robba; Metcalfe, Peter; Walker, Amy; Holloway, Lois.
Afiliación
  • Carr ME; Centre for Medical and Radiation Physics, University of Wollongong, Wollongong, Australia.
  • Keenan KE; Ingham Institute for Applied Medical Research, Liverpool, Australia.
  • Rai R; Liverpool and Macarthur Cancer Therapy Centres, Sydney, Australia.
  • Metcalfe P; National Institute of Standards and Technology, Boulder, Colorado, USA.
  • Walker A; Ingham Institute for Applied Medical Research, Liverpool, Australia.
  • Holloway L; Liverpool and Macarthur Cancer Therapy Centres, Sydney, Australia.
J Appl Clin Med Phys ; 22(11): 143-150, 2021 Nov.
Article en En | MEDLINE | ID: mdl-34562341
ABSTRACT

PURPOSE:

To determine baseline accuracy and reproducibility of T1 and T2 relaxation times over 12 months on a dedicated radiotherapy MRI scanner.

METHODS:

An International Society of Magnetic Resonance in Medicine/National Institute of Standards and Technology (ISMRM/NIST) System Phantom was scanned monthly on a 3T MRI scanner for 1 year. T1 was measured using inversion recovery (T1 -IR) and variable flip angle (T1 -VFA) sequences and T2 was measured using a multi-echo spin echo (T2 -SE) sequence. For each vial in the phantom, accuracy errors (%bias) were determined by the relative differences in measured T1 and T2 times compared to reference values. Reproducibility was measured by the coefficient of variation (CV) of T1 and T2 measurements across monthly scans. Accuracy and reproducibility were mainly assessed on vials with relaxation times expected to be in physiological ranges at 3T.

RESULTS:

A strong linear correlation between measured and reference relaxation times was found for all sequences tested (R2  > 0.997). Baseline bias (and CV[%]) for T1 -IR, T1 -VFA and T2 -SE sequences were +2.0% (2.1), +6.5% (4.2), and +8.5% (1.9), respectively.

CONCLUSIONS:

The accuracy and reproducibility of T1 and T2 on the scanner were considered sufficient for the sequences tested. No longitudinal trends of variation were deduced, suggesting less frequent measurements are required following the establishment of baselines.
Asunto(s)
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2021 Tipo del documento: Article País de afiliación: Australia

Texto completo: 1 Banco de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética Límite: Humans Idioma: En Revista: J Appl Clin Med Phys Asunto de la revista: BIOFISICA Año: 2021 Tipo del documento: Article País de afiliación: Australia