ABSTRACT
PURPOSE:
Highly localized impedance (LI) measurements during atrial fibrillation (AF) ablation have emerged as a viable real-time indicator of tissue characteristics and the consequent durability of the lesions created. We investigated the impact of catheter-tissue contact force (CF) on LI behavior during pulmonary vein isolation (PVI).METHODS:
Forty-five consecutive patients of the CHARISMA registry undergoing de novo AF radiofrequency (RF) catheter ablation with a novel open-irrigated-tip catheter endowed with CF and LI measurement capabilities (Stablepoint™ catheter, Boston Scientific) were included.RESULTS:
A total of 2895 point-by-point RF applications were analyzed (RF delivery time (DT) = 8.7±4s, CF = 13 ±±8 g, LI drop = 23 ±±7 Ω). All PVs were successfully isolated in an overall procedure time of 118 ±±34 min (fluoroscopy time = 13 ±±8 min). The magnitude of LI drop weakly correlated with CF (r = 0.13, 95% confidence interval (CI) 0.09 to 0.16, p < 0.0001), whereas both CF and LI drop inversely correlated with DT (r = -0.26, 95%CI -0.29 to -0.22, p < 0.0001 for CF; r = -0.36, 95%CI -0.39 to -0.33, p < 0.0001 for LI). For each 10 g of CF, LI drop markedly increased from 22.4 ± 7 Ω to 24.0 ± 8 Ω at 5 to 25 g CF intervals (5-14 g of CF vs 15-24 g of CF, p < 0.0001), whereas it showed smooth transition over 25 g (24.8 ± 7Ω at ≥ 25 g CF intervals, p = 0.0606 vs 15-24 g of CF). No major complications occurred during the procedures or within 30 days.CONCLUSIONS:
CF significantly affects LI drop and probable consequent lesion formation during RF PVI. The benefit of higher contact (> 25 g) between the catheter and the tissue appears to have less impact on LI drop. TRIAL REGISTRATION Catheter Ablation of Arrhythmias With High Density Mapping System in the Real World Practice (CHARISMA). URL http//clinicaltrials.gov/ Identifier NCT03793998.Palabras clave
Texto completo:
1
Banco de datos:
MEDLINE
Asunto principal:
Venas Pulmonares
/
Fibrilación Atrial
/
Ablación por Catéter
Tipo de estudio:
Prognostic_studies
Límite:
Humans
Idioma:
En
Revista:
J Interv Card Electrophysiol
Asunto de la revista:
CARDIOLOGIA
Año:
2022
Tipo del documento:
Article
País de afiliación:
Italia