Your browser doesn't support javascript.
loading
Outcome of MRI in Patients with Nonconditional Devices with Mismatch between Manufacturer of Leads and Generator.
Okasha, Osama; Saeed, Ibrahim M; Gupta, Sanjaya K.
Afiliación
  • Okasha O; Department of Medicine, Saint Luke's Health System, Kansas City, Mo (O.O.); INOVA Heart and Vascular and Virginia Heart, Falls Church, Va (I.M.S.); Division of Cardiology, Saint Luke's Mid America Heart Institute, 4401 Wornall Rd, Kansas City, MO 64111 (S.K.G.); and University of Missouri Kansas City School of Medicine, Kansas City, Mo (S.K.G.).
  • Saeed IM; Department of Medicine, Saint Luke's Health System, Kansas City, Mo (O.O.); INOVA Heart and Vascular and Virginia Heart, Falls Church, Va (I.M.S.); Division of Cardiology, Saint Luke's Mid America Heart Institute, 4401 Wornall Rd, Kansas City, MO 64111 (S.K.G.); and University of Missouri Kansas City School of Medicine, Kansas City, Mo (S.K.G.).
  • Gupta SK; Department of Medicine, Saint Luke's Health System, Kansas City, Mo (O.O.); INOVA Heart and Vascular and Virginia Heart, Falls Church, Va (I.M.S.); Division of Cardiology, Saint Luke's Mid America Heart Institute, 4401 Wornall Rd, Kansas City, MO 64111 (S.K.G.); and University of Missouri Kansas City School of Medicine, Kansas City, Mo (S.K.G.).
Radiol Cardiothorac Imaging ; 4(3): e220014, 2022 Jun.
Article en En | MEDLINE | ID: mdl-35833162
ABSTRACT
In a series of 35 MRI examinations with non-MRI-conditional devices with a mismatch between the manufacturer of the device generators and leads, there were no adverse events.
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Revista: Radiol Cardiothorac Imaging Año: 2022 Tipo del documento: Article