Artifact reduction in low and ultra-low dose chest computed tomography for patients with pacemaker: A phantom study.
Watanabe, S; Urikura, A; Ohashi, K; Kitera, N; Tsuchiya, T; Kasai, H; Kawai, T; Hiwatashi, A.
Afiliación
  • Watanabe S; Department of Radiology, Nagoya City University Hospital, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, Aichi 467-0001, Japan. Electronic address: s_w_zeal@yahoo.co.jp.
  • Urikura A; Department of Radiological Technology, Radiological Diagnosis, National Cancer Center Hospital, 5-1-1 Tsukiji, Chuo-ku, Tokyo, 104-0045, Japan. Electronic address: aurikura@ncc.go.jp.
  • Ohashi K; Department of Radiology, Nagoya City University Midori Municipal Hospital, 1-77 Shiomigaoka, Midori-ku, Nagoya, Aichi, 458-0037, Japan; Department of Radiology, Nagoya City University Graduate School of Medical Sciences, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, Aichi 467-0001, Japan. Electronic ad
  • Kitera N; Department of Radiology, Nagoya City University Hospital, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, Aichi 467-0001, Japan. Electronic address: n.kitera@gmail.com.
  • Tsuchiya T; Department of Radiology, Nagoya City University Hospital, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, Aichi 467-0001, Japan. Electronic address: ratsuchi@med.nagoya-cu.ac.jp.
  • Kasai H; Department of Radiology, Nagoya City University Hospital, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, Aichi 467-0001, Japan. Electronic address: rahkasai@med.nagoya-cu.ac.jp.
  • Kawai T; Department of Radiology, Nagoya City University Midori Municipal Hospital, 1-77 Shiomigaoka, Midori-ku, Nagoya, Aichi, 458-0037, Japan; Department of Radiology, Nagoya City University Graduate School of Medical Sciences, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, Aichi 467-0001, Japan. Electronic ad
  • Hiwatashi A; Department of Radiology, Nagoya City University Graduate School of Medical Sciences, 1 Kawasumi, Mizuho-cho, Mizuho-ku, Nagoya, Aichi 467-0001, Japan. Electronic address: ahiwata@med.nagoya-cu.ac.jp.
Radiography (Lond) ; 30(3): 770-775, 2024 05.
Article en En | MEDLINE | ID: mdl-38460224
ABSTRACT

INTRODUCTION:

Implanted pacemakers (PM) would decrease the detection of lung nodules in chest computed tomography (CT) due to the metal artifact. This study aimed to explore the computer-aided diagnosis (CAD) detectability of pulmonary nodules for the patients implanted with PMs in low- and ultra-low-dose chest CT screening.

METHODS:

Four different sizes of artificial nodules were placed in an anthropomorphic chest phantom with two alternative diameters utilized. A commercially available PM was placed on the surface of the left chest wall of the phantom. The image acquisitions were performed with 120 kV and 150 kV with a dedicated selective photon shield made of tin filter (Sn150 kV) at low- and ultra-low- radiation doses (1.0 and 0.5 mGy of volume CT dose index), and reconstructed with and without Iterative Metal Artifact Reduction (iMAR, Siemens Healthineers, Erlangen, Germany). The relative artifact index (AIr) was calculated as an index of metal artifacts, and the nodule detectability was evaluated with a CAD system.

RESULTS:

Sn150 kV reduced AIr in all acquisitions when comparing 120 kV and Sn150 kV. Although PM reduced the detectability of nodules, Sn150 kV showed higher detectability compared to 120 kV. The use of iMAR showed inconsistent results in nodule detectability.

CONCLUSION:

Sn150 kV reduced PM-induced metal artifacts and improved nodule detectability with CAD compared to 120 kV acquisition in many conditions including low and ultra-low doses and large phantoms, but iMAR did not improve the detectability. IMPLICATIONS FOR PRACTICE Based on the results of the current phantom study, low and ultra-low dose with Sn150 kV acquisition reduced PM-induced metal artifacts and improved nodule detectability.