Your browser doesn't support javascript.
loading
Treatment of intracranial aneurysms with large-diameter (≥5.5 mm) Derivo Embolization Devices, with particular focus on 7 and 8 mm diameter devices.
Goertz, Lukas; Zopfs, David; Kottlors, Jonathan; Pennig, Lenhard; Schob, Stefan; Schlamann, Marc; Kabbasch, Christoph.
Afiliación
  • Goertz L; Faculty of Medicine and University Hospital, Department of Radiology and Neuroradiology, University of Cologne, Cologne, Germany.
  • Zopfs D; Faculty of Medicine and University Hospital, Department of Radiology and Neuroradiology, University of Cologne, Cologne, Germany.
  • Kottlors J; Faculty of Medicine and University Hospital, Department of Radiology and Neuroradiology, University of Cologne, Cologne, Germany.
  • Pennig L; Faculty of Medicine and University Hospital, Department of Radiology and Neuroradiology, University of Cologne, Cologne, Germany.
  • Schob S; Department of Radiology and Neuroradiology, University Hospital Halle, Halle, Germany.
  • Schlamann M; Faculty of Medicine and University Hospital, Department of Radiology and Neuroradiology, University of Cologne, Cologne, Germany.
  • Kabbasch C; Faculty of Medicine and University Hospital, Department of Radiology and Neuroradiology, University of Cologne, Cologne, Germany.
Interv Neuroradiol ; : 15910199241248479, 2024 May 05.
Article en En | MEDLINE | ID: mdl-38706147
ABSTRACT

OBJECTIVE:

There are few studies on flow diverters with diameters >5 mm. We present our preliminary experience with the 5.5-8 mm Derivo Embolization Device (DED) implants for the treatment of cerebral aneurysms.

METHODS:

A consecutive series of 26 patients (median age 60 years) treated for 32 aneurysms in 26 procedures was retrospectively analyzed for procedural characteristics, complications, and mid-term angiographic results.

RESULTS:

The median aneurysm size was 10.5 mm, 2 of 30 (6%) aneurysms were ruptured and 9 (28%) had a fusiform or dissecting morphology. DED implantation was performed in the internal carotid artery in 18 of 26 (69%) procedures and in the vertebrobasilar artery in 8 (31%). Device deployment failed in 1 (4%) procedure. The 7 and 8 mm implants were successfully deployed in 5 cases. Additional balloon angioplasty or stent implantation was performed in 3 (12%) cases to improve wall apposition. Complications included 1 (4%) major stroke and 2 (8%) minor strokes. Angiographic follow up at a mean of 6 months showed complete occlusion in 8 of30 (27%) aneurysms and favorable occlusion in 14 (47%).

CONCLUSIONS:

The use of large diameter DEDs was safe and feasible. The mid-term occlusion rates are acceptable considering the complex subset of aneurysms studied. Further studies are warranted to define the indications for large-diameter DEDs and to evaluate their long-term efficacy.
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Revista: Interv Neuroradiol Asunto de la revista: NEUROLOGIA / RADIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Revista: Interv Neuroradiol Asunto de la revista: NEUROLOGIA / RADIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Alemania