Size and fluorescence calibrated imaging flow cytometry: From arbitrary to standard units.
Woud, Wouter W; Pugsley, Haley R; Bettin, Britta A; Varga, Zoltán; van der Pol, Edwin.
Afiliación
  • Woud WW; Erasmus MC Transplant Institute, Department of Internal Medicine, University Medical Center Rotterdam, Rotterdam, The Netherlands.
  • Pugsley HR; Application Cytometry, Cytek Biosciences, Inc, Seattle, Washington, USA.
  • Bettin BA; Department of Biomedical Engineering and Physics, Amsterdam University Medical Center location University of Amsterdam, Amsterdam, The Netherlands.
  • Varga Z; Department of Clinical Chemistry, Amsterdam University Medical Center location University of Amsterdam, Amsterdam, The Netherlands.
  • van der Pol E; Amsterdam Vesicle Center, Amsterdam University Medical Center location University of Amsterdam, Amsterdam, The Netherlands.
Cytometry A ; 2024 Sep 05.
Article en En | MEDLINE | ID: mdl-39238272
ABSTRACT
Imaging flow cytometry (IFCM) is a technique that can detect, size, and phenotype extracellular vesicles (EVs) at high throughput (thousands/minute) in complex biofluids without prior EV isolation. However, the generated signals are expressed in arbitrary units, which hinders data interpretation and comparison of measurement results between instruments and institutes. While fluorescence calibration can be readily achieved, calibration of side scatter (SSC) signals presents an ongoing challenge for IFCM. Here, we present an approach to relate the SSC signals to particle size for IFCM, and perform a comparability study between three different IFCMs using a plasma EV test sample (PEVTES). SSC signals for different sizes of polystyrene (PS) and hollow organosilica beads (HOBs) were acquired with a 405 nm 120 mW laser without a notch filter before detection. Mie theory was applied to relate scatter signals to particle size. Fluorescence calibration was accomplished with 2 µm phycoerythrin (PE) and allophycocyanin (APC) MESF beads. Size and fluorescence calibration was performed for three IFCMs in two laboratories. CD235a-PE and CD61-APC stained PEVTES were used as EV-containing samples. EV concentrations were compared between instruments within a size range of 100-1000 nm and a fluorescence intensity range of 3-10,000 MESF. 81 nm PS beads could be readily discerned from background based on their SSC signals. Fitting of the obtained PS bead SSC signals with Mie theory resulted in a coefficient of determination >0.99 between theory and data for all three IFCMs. 216 nm HOBs were detected with all instruments, and confirmed the sensitivity to detect EVs by SSC. The lower limit of detection regarding EV-size for this study was determined to be ~100 nm for all instruments. Size and fluorescence calibration of IFCM data increased cross-instrument data comparability with the coefficient of variation decreasing from 33% to 21%. Here we demonstrate - for the first time - scatter calibration of an IFCM using the 405 nm laser. The quality of the scatter-to-diameter relation and scatter sensitivity of the IFCMs are similar to the most sensitive commercially available flow cytometers. This development will support the reliability of EV research with IFCM by providing robust standardization and reproducibility, which are pre-requisites for understanding the biological significance of EVs.
Palabras clave

Texto completo: 1 Banco de datos: MEDLINE Idioma: En Revista: Cytometry A / Cytometry, Part A / Cytometry. Part A Año: 2024 Tipo del documento: Article País de afiliación: Países Bajos