Your browser doesn't support javascript.
loading
Pitfalls in the somatomedin-C radioimmunoassay.
J Clin Endocrinol Metab ; 54(5): 1084-6, 1982 May.
Article en En | MEDLINE | ID: mdl-7199537
ABSTRACT
We have investigated the effects of various modifications on the quantitation of Somatomedin-C (Sm-C) by the Sm-C radio immunoassay. We have confirmed that the assay quantitatively measures Sm-C in serum at plasma samples when borosilicate tubes and protamine free buffers are used. Polystyrene tubes and/or protamine containing buffers interfere with the quantitative measurement of the hormone, apparently by modifying the binding of Sm-C to its binding protein. The implication of these modifications on the clinical use of the Sm-C RIA are discussed.
Asunto(s)
Buscar en Google
Banco de datos: MEDLINE Asunto principal: Somatomedinas Límite: Humans Idioma: En Revista: J Clin Endocrinol Metab Año: 1982 Tipo del documento: Article
Buscar en Google