Design of a testing system for antithrombotic pressure circulatory equipment / 生物医学工程学杂志
Article en Zh | WPRIM | ID: wpr-259671
Biblioteca responsable: WPRO
ABSTRACT
A new type of testing system used for antithrombotic pressure circulatory equipment has been developed, which realized a new method for the calibration of pressure sensor. Multi-path control and acquisition functions are achieved by this method based on human-computer interaction testing system. The precision of pressure sensor is ob tained by polynomial fitting for each test point using linear interpolation method. The result showed that the precision test of pressure sensor could be realized easily and efficiently, using the developed testing system, and the parameters of pressure sensor could be calibrated effectively, so that it could be accurately used in the antithrombotic pressure circulatory equipment. The developed testing system has a prosperous future in the aspects of promotion and application.
Asunto(s)
Texto completo: 1 Banco de datos: WPRIM Asunto principal: Presión / Trombosis / Algoritmos / Calibración / Diagnóstico / Diseño de Equipo / Monitoreo Fisiológico Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: Zh Revista: Journal of Biomedical Engineering Año: 2014 Tipo del documento: Article
Texto completo: 1 Banco de datos: WPRIM Asunto principal: Presión / Trombosis / Algoritmos / Calibración / Diagnóstico / Diseño de Equipo / Monitoreo Fisiológico Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: Zh Revista: Journal of Biomedical Engineering Año: 2014 Tipo del documento: Article