ABSTRACT
This paper introduces the KRISS-Rn4, a high-sensitivity radon monitor with four detection cells, installed within a walk-in type radon calibration chamber at KRISS. The KRISS-Rn4 exhibits enhanced energy resolution through channel-by-channel signal processing and data acquisition. Results reveal that it achieves lower statistical fluctuations and faster response times in monitoring test atmospheres compared to commercial devices.
ABSTRACT
A 4πß(PPC)-γ coincidence system has been made at KRISS based on a digital DAQ. 60Co sources were measured to verify the system. The maximum detection efficiency for beta particles was estimated to be 96.7 %. Massic activities for sample sources had 0.005 % of the sample variability error, which was well within the expanded standard uncertainty of 0.54 % (k = 2).