1.
Magn Reson Med ; 91(2): 640-648, 2024 Feb.
Article in English | MEDLINE | ID: mdl-37753628

ABSTRACT

PURPOSE: To demonstrate the technical feasibility and the value of ultrahigh-performance gradient in imaging the prostate in a 3T MRI system. METHODS: In this local institutional review board-approved study, prostate MRI was performed on 4 healthy men. Each subject was scanned in a prototype 3T MRI system with a 42-cm inner-diameter gradient coil that achieves a maximum gradient amplitude of 200 mT/m and slew rate of 500 T/m/s. PI-RADS V2.1-compliant axial T2 -weighted anatomical imaging and single-shot echo planar DWI at standard gradient of 70 mT/m and 150 T/m/s were obtained, followed by DWI at maximum performance (i.e., 200 mT/m and 500 T/m/s). In comparison to state-of-the-art clinical whole-body MRI systems, the high slew rate improved echo spacing from 1020 to 596 µs and, together with a high gradient amplitude for diffusion encoding, TE was reduced from 55 to 36 ms. RESULTS: In all 4 subjects (waist circumference = 81-91 cm, age = 45-65 years), no peripheral nerve stimulation sensation was reported during DWI. Reduced image distortion in the posterior peripheral zone prostate gland and higher signal intensity, such as in the surrounding muscle of high-gradient DWI, were noted. CONCLUSION: Human prostate MRI at simultaneously high gradient amplitude of 200 mT/m and slew rate of 500 T/m/s is feasible, demonstrating that improved gradient performance can address image distortion and T2 decay-induced SNR issues for in vivo prostate imaging.


Subject(s)
Magnetic Resonance Imaging , Prostatic Neoplasms , Male , Humans , Middle Aged , Aged , Prostate/diagnostic imaging , Feasibility Studies , Diffusion Magnetic Resonance Imaging/methods , Echo-Planar Imaging/methods , Prostatic Neoplasms/diagnostic imaging , Reproducibility of Results
2.
Magn Reson Med ; 89(1): 262-275, 2023 01.
Article in English | MEDLINE | ID: mdl-36129000

ABSTRACT

PURPOSE: Asymmetric gradient coils introduce zeroth- and first-order concomitant field terms, in addition to higher-order terms common to both asymmetric and symmetric gradients. Salient to compensation strategies is the accurate calibration of the concomitant field spatial offset parameters for asymmetric coils. A method that allows for one-time calibration of the offset parameters is described. THEORY AND METHODS: A modified phase contrast pulse sequence with single-sided bipolar flow encoding is proposed to calibrate the offsets for asymmetric, transverse gradient coils. By fitting the measured phase offsets to different gradient amplitudes, the spatial offsets were calculated by fitting the phase variation. This was used for calibrating real-time pre-emphasis compensation of the zeroth- and first-order concomitant fields. RESULTS: Image quality improvement with the proposed corrections was demonstrated in phantom and healthy volunteers with non-Cartesian and Cartesian trajectory acquisitions. Concomitant field compensation using the calibrated offsets resulted in a residual phase error <3% at the highest gradient amplitude and demonstrated substantial reduction of image blur and slice position/selection artifacts. CONCLUSIONS: The proposed implementation provides an accurate method for calibrating spatial offsets that can be used for real-time concomitant field compensation of zeroth and first-order terms, substantially reducing artifacts without retrospective correction or sequence specific waveform modifications.


Subject(s)
Image Processing, Computer-Assisted , Magnetic Resonance Imaging , Humans , Image Processing, Computer-Assisted/methods , Calibration , Retrospective Studies , Magnetic Resonance Imaging/methods , Artifacts , Phantoms, Imaging
3.
Magn Reson Med ; 84(2): 950-965, 2020 08.
Article in English | MEDLINE | ID: mdl-32011027

ABSTRACT

PURPOSE: We investigate the importance of high gradient-amplitude and high slew-rate on oscillating gradient spin echo (OGSE) diffusion imaging for human brain imaging and evaluate human brain imaging with OGSE on the MAGNUS head-gradient insert (200 mT/m amplitude and 500 T/m/s slew rate). METHODS: Simulations with cosine-modulated and trapezoidal-cosine OGSE at various gradient amplitudes and slew rates were performed. Six healthy subjects were imaged with the MAGNUS gradient at 3T with OGSE at frequencies up to 100 Hz and b = 450 s/mm2 . Comparisons were made against standard pulsed gradient spin echo (PGSE) diffusion in vivo and in an isotropic diffusion phantom. RESULTS: Simulations show that to achieve high frequency and b-value simultaneously for OGSE, high gradient amplitude, high slew rates, and high peripheral nerve stimulation limits are required. A strong linear trend for increased diffusivity (mean: 8-19%, radial: 9-27%, parallel: 8-15%) was observed in normal white matter with OGSE (20 Hz to 100 Hz) as compared to PGSE. Linear fitting to frequency provided excellent correlation, and using a short-range disorder model provided radial long-term diffusivities of D∞,MD = 911 ± 72 µm2 /s, D∞,PD = 1519 ± 164 µm2 /s, and D∞,RD = 640 ± 111 µm2 /s and correlation lengths of lc,MD = 0.802 ± 0.156 µm, lc,PD = 0.837 ± 0.172 µm, and lc,RD = 0.780 ± 0.174 µm. Diffusivity changes with OGSE frequency were negligible in the phantom, as expected. CONCLUSION: The high gradient amplitude, high slew rate, and high peripheral nerve stimulation thresholds of the MAGNUS head-gradient enables OGSE acquisition for in vivo human brain imaging.


Subject(s)
Brain , Diffusion Magnetic Resonance Imaging , Brain/diagnostic imaging , Diffusion , Humans , Neuroimaging , Phantoms, Imaging
4.
Magn Reson Med ; 83(1): 352-366, 2020 01.
Article in English | MEDLINE | ID: mdl-31385628

ABSTRACT

PURPOSE: To establish peripheral nerve stimulation (PNS) thresholds for an ultra-high performance magnetic field gradient subsystem (simultaneous 200-mT/m gradient amplitude and 500-T/m/s gradient slew rate; 1 MVA per axis [MAGNUS]) designed for neuroimaging with asymmetric transverse gradients and 42-cm inner diameter, and to determine PNS threshold dependencies on gender, age, patient positioning within the gradient subsystem, and anatomical landmarks. METHODS: The MAGNUS head gradient was installed in a whole-body 3T scanner with a custom 16-rung bird-cage transmit/receive RF coil compatible with phased-array receiver brain coils. Twenty adult subjects (10 male, mean ± SD age = 40.4 ± 11.1 years) underwent the imaging and PNS study. The tests were repeated by displacing subject positions by 2-4 cm in the superior-inferior and anterior-posterior directions. RESULTS: The x-axis (left-right) yielded mostly facial stimulation, with mean ΔGmin = 111 ± 6 mT/m, chronaxie = 766 ± 76 µsec. The z-axis (superior-inferior) yielded mostly chest/shoulder stimulation (123 ± 7 mT/m, 620 ± 62 µsec). Y-axis (anterior-posterior) stimulation was negligible. X-axis and z-axis thresholds tended to increase with age, and there was negligible dependency with gender. Translation in the inferior and posterior directions tended to increase the x-axis and z-axis thresholds, respectively. Electric field simulations showed good agreement with the PNS results. Imaging at MAGNUS gradient performance with increased PNS threshold provided a 35% reduction in noise-to-diffusion contrast as compared with whole-body performance (80 mT/m gradient amplitude, 200 T/m/sec gradient slew rate). CONCLUSION: The PNS threshold of MAGNUS is significantly higher than that for whole-body gradients, which allows for diffusion gradients with short rise times (under 1 msec), important for interrogating brain microstructure length scales.


Subject(s)
Brain/diagnostic imaging , Electric Stimulation , Magnetic Fields , Neuroimaging/instrumentation , Neuroimaging/methods , Peripheral Nerves/diagnostic imaging , Peripheral Nervous System/diagnostic imaging , Adult , Algorithms , Equipment Design , Female , Humans , Image Processing, Computer-Assisted , Magnetic Resonance Imaging , Male , Middle Aged , Peripheral Nerves/physiology , Phantoms, Imaging , Reproducibility of Results , Whole Body Imaging
5.
Magn Reson Med ; 83(6): 2356-2369, 2020 06.
Article in English | MEDLINE | ID: mdl-31763726

ABSTRACT

PURPOSE: To develop a highly efficient magnetic field gradient coil for head imaging that achieves 200 mT/m and 500 T/m/s on each axis using a standard 1 MVA gradient driver in clinical whole-body 3.0T MR magnet. METHODS: A 42-cm inner diameter head-gradient used the available 89- to 91-cm warm bore space in a whole-body 3.0T magnet by increasing the radial separation between the primary and the shield coil windings to 18.6 cm. This required the removal of the standard whole-body gradient and radiofrequency coils. To achieve a coil efficiency ~4× that of whole-body gradients, a double-layer primary coil design with asymmetric x-y axes, and symmetric z-axis was used. The use of all-hollow conductor with direct fluid cooling of the gradient coil enabled ≥50 kW of total heat dissipation. RESULTS: This design achieved a coil efficiency of 0.32 mT/m/A, allowing 200 mT/m and 500 T/m/s for a 620 A/1500 V driver. The gradient coil yielded substantially reduced echo spacing, and minimum repetition time and echo time. In high b = 10,000 s/mm2 diffusion, echo time (TE) < 50 ms was achieved (>50% reduction compared with whole-body gradients). The gradient coil passed the American College of Radiology tests for gradient linearity and distortion, and met acoustic requirements for nonsignificant risk operation. CONCLUSIONS: Ultra-high gradient coil performance was achieved for head imaging without substantial increases in gradient driver power in a whole-body 3.0T magnet after removing the standard gradient coil. As such, any clinical whole-body 3.0T MR system could be upgraded with 3-4× improvement in gradient performance for brain imaging.


Subject(s)
Brain , Magnetic Resonance Imaging , Acoustics , Brain/diagnostic imaging , Equipment Design , Head/diagnostic imaging , Humans , Magnetic Fields
6.
Magn Reson Med ; 81(2): 803-810, 2019 02.
Article in English | MEDLINE | ID: mdl-30325052

ABSTRACT

MR-based electrical properties tomography converts the MRI transmit/receive RF field measurements to tissue electrical property maps through dedicated reconstruction algorithms. Recent reports showed that despite limitations, electrical properties tomography holds promise for generating additional contrast for tumor detection and patient-specific modeling of tissue-RF field interactions. This review summarizes the available tissue electrical property contrasts and compares them with the capabilities of the most commonly used electrical properties tomography reconstruction method. Future directions and prospects of clinical translation are discussed.


Subject(s)
Contrast Media/chemistry , Electric Conductivity , Electric Impedance , Image Processing, Computer-Assisted/methods , Magnetic Resonance Imaging , Algorithms , Brain/diagnostic imaging , Breast/diagnostic imaging , Female , Gray Matter/diagnostic imaging , Healthy Volunteers , Humans , Patient-Specific Modeling , Phantoms, Imaging , Reproducibility of Results , Tomography , White Matter/diagnostic imaging
7.
Magn Reson Med ; 80(5): 2232-2245, 2018 11.
Article in English | MEDLINE | ID: mdl-29536587

ABSTRACT

PURPOSE: To build and evaluate a small-footprint, lightweight, high-performance 3T MRI scanner for advanced brain imaging with image quality that is equal to or better than conventional whole-body clinical 3T MRI scanners, while achieving substantial reductions in installation costs. METHODS: A conduction-cooled magnet was developed that uses less than 12 liters of liquid helium in a gas-charged sealed system, and standard NbTi wire, and weighs approximately 2000 kg. A 42-cm inner-diameter gradient coil with asymmetric transverse axes was developed to provide patient access for head and extremity exams, while minimizing magnet-gradient interactions that adversely affect image quality. The gradient coil was designed to achieve simultaneous operation of 80-mT/m peak gradient amplitude at a slew rate of 700 T/m/s on each gradient axis using readily available 1-MVA gradient drivers. RESULTS: In a comparison of anatomical imaging in 16 patients using T2 -weighted 3D fluid-attenuated inversion recovery (FLAIR) between the compact 3T and whole-body 3T, image quality was assessed as equivalent to or better across several metrics. The ability to fully use a high slew rate of 700 T/m/s simultaneously with 80-mT/m maximum gradient amplitude resulted in improvements in image quality across EPI, DWI, and anatomical imaging of the brain. CONCLUSIONS: The compact 3T MRI system has been in continuous operation at the Mayo Clinic since March 2016. To date, over 200 patient studies have been completed, including 96 comparison studies with a clinical 3T whole-body MRI. The increased gradient performance has reliably resulted in consistently improved image quality.


Subject(s)
Magnetic Resonance Imaging/instrumentation , Whole Body Imaging/instrumentation , Brain/diagnostic imaging , Equipment Design , Female , Humans , Imaging, Three-Dimensional , Magnets , Male , Phantoms, Imaging , Signal-To-Noise Ratio
8.
Magn Reson Imaging ; 103: 109-118, 2023 11.
Article in English | MEDLINE | ID: mdl-37468020

ABSTRACT

Access to high-quality MR exams is severely limited for patients with some implanted devices due to labeled MR safety conditions, but small-bore systems can overcome this limitation. For example, a compact 3 T MR scanner (C3T) with high-performance gradients can acquire exams of the head, extremities, and infants. Because of its reduced bore size and the patient being advanced only partially into the bore, the associated electromagnetic (EM) fields drop off rapidly caudal to the head, compared to whole-body systems. Therefore, some patients with MR conditional implanted devices can safely receive 3 T brain exams on the C3T using its strong gradients and a multiple-channel receive coil, while a corresponding exam on whole-body MR is precluded. The purpose of this study is to evaluate the performance of a small-bore scanner for subjects with MR conditional spinal or sacral nerve stimulators, or abandoned cardiac implantable electronic device (CIED) leads. The spatial dependence of specific absorption rate (SAR) on the C3T was compared to whole-body scanners. A device assessment tool was developed and applied to evaluate MR safety individually on the C3T for 12 subjects with implanted devices or abandoned CIED leads. Once MR safety was established, the subjects received a C3T brain exam along with their clinical, 1.5 T exam. The resulting images were graded by three board-certified neuroradiologists. The C3T exams were well-tolerated with no adverse events, and significantly outperformed the whole-body 1.5 T exams in terms of overall image quality.


Subject(s)
Magnetic Resonance Imaging , Prostheses and Implants , Humans , Magnetic Resonance Imaging/methods , Brain/diagnostic imaging , Heart , Head