Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 16 de 16
Filter
Add more filters

Publication year range
1.
Artif Organs ; 47(2): 342-351, 2023 Feb.
Article in English | MEDLINE | ID: mdl-36134430

ABSTRACT

INTRODUCTION: In vitro hemolysis testing is an essential method for assessing the hemolytic potential of blood pumps, but has poor reproducibility. Further investigations are needed to determine the sources and extent of variability and to find a practical way to reduce the variation. METHODS: A small volume blood circulating loop driven by a Centrimag pump was established to provide relatively higher hemolysis readouts within a short run time and to be able to sequentially perform multiple repeated hemolysis tests in a working day. RESULTS: The repeatability with this system was demonstrated as the %RSD at 4.3% for the NIH or MIH from three repeated tests using the same blood. The bovine blood from different randomly selected donors was tested and gave more than a two-fold difference in NIH results (0.077 vs. 0.032 g/100 L) under the same testing conditions and same pump. This wide variation in hemolysis using bovine blood from different donors happened repeatedly. More importantly, it was observed that the difference in hemolysis test results using the blood drawn from the same donor on multiple days was narrow although the native hematocrits varied. The %RSD of NIH values obtained on five different days were 6.8%, 8.4%, 11.5%, and 7.8% for donor-specific blood from donors 1 to 4, respectively. CONCLUSION: The study results indicate that the mechanical stress-induced hemolysis behavior is donor-dependent. It has been also demonstrated that the reproducibility of in vitro hemolysis testing can be improved when the blood drawn from same donor is used.


Subject(s)
Assisted Circulation , Heart-Assist Devices , Animals , Cattle , Hemolysis , Stress, Mechanical , Reproducibility of Results , Hematocrit
2.
ACS Biomater Sci Eng ; 9(5): 2277-2291, 2023 05 08.
Article in English | MEDLINE | ID: mdl-37094361

ABSTRACT

The chemical characterization of extractables and leachables (E&Ls) is an important aspect of biosafety and biocompatibility assessment in medical device industry. The advent of the body-contact use of medical devices in patient treatment has introduced a potential source for extractables and leachables as these medical devices are comprised of various polymeric materials. Several industry working groups, the FDA and USP, have recognized the guidance for chemical characterizations and nontargeted analysis of medical device extracts, such as ISO 10993-18:2020. The MS application of nontargeted analysis has played a critical role in understanding the E&Ls from medical device extracts. However, there have been very few reports about the MS based workflow with nontargeted analysis for medical device extracts and there is little guidance about the exact methodologies which should be used, even though there is an urgent need for a clearly defined process for the identification of medical device extracts. In this study, we demonstrated an analytical LC/MS (liquid chromatography/mass spectrometry) workflow using high resolution Exploris120 Orbitrap instrument for data acquisition and Compound Discoverer 3.3 intelligent software for data processing to profile the polymer related E&Ls from a balloon dilation catheter device extracted with 40% ethanol. An E&L ID workflow combining LC separation, data-informed MS acquisition strategy, MS information mining (including adduct ions, MS information from both electrospray ionization (ESI) (+) and ESI (-), in-source fragmentation, common fragment ions (CFIs), common neutral losses (CNLs), and in silico MS simulation was described with intelligent software processing and manual data interpretation. The workflow developed in this study was proven to be effective to provide a comprehensive profile of polymer related degradation products, polymer impurities and additives including surfactants, UV curing agent, antioxidants, and plasticizers for the device analyzed. The classification of E&L compounds using CFIs and CNLs was very effective to facilitate the identification of polymer related impurities and extract the polymer related impurities with common structures in a large data result set.


Subject(s)
Complex Mixtures , Polymers , Humans , Workflow , Mass Spectrometry , Chromatography, Liquid
3.
ACS Omega ; 7(51): 48291-48300, 2022 Dec 27.
Article in English | MEDLINE | ID: mdl-36591149

ABSTRACT

Extractables or leachables of biomaterials or residues of additives used in the manufacturing process that are potentially released from a medical device may have an adverse effect on a patient. Chemical characterization of leachable chemicals and degradation products in a medical device is an important aspect of its overall biocompatibility assessment process, which helps to ensure that the therapeutical benefits exceed the potential biological risks associated with the use of the device or its components or materials. By evaluating the types and amounts of chemicals that may migrate from a device to a patient during clinical use, potential toxicological risks can be assessed. A semipolar solvent, 40% ethanol in water (v/v), an appropriate surrogate for blood and blood related substances, was used as an extraction medium to mimic the body fluid in contact with a medical device. The extraction was conducted at 37 °C for 24 h for limited exposure medical devices per ISO 10993-12:2021. From gas chromatography-mass spectrometry (GC-MS) and liquid chromatography-mass spectrometry (LC-MS) analysis, leachable chemicals of polylactams, linear polyamides, cyclic polytetramethylene ether (PTME), poly(tetramethylene ether) glycol (PTMEG), cyclic and linear poly(tetramethylene ether) glycol adipate (PTMEGA), cyclic and linear poly(tetramethylene ether) glycol adipamide (PTMEGAA) were structurally elucidated. The workflow presented in this study was proven to be a successful approach for rapid extractable and leachable profiling and identification with confidence.

4.
Circulation ; 122(22): 2288-300, 2010 Nov 30.
Article in English | MEDLINE | ID: mdl-20975003

ABSTRACT

BACKGROUND: With the use of optical coherence tomography (OCT), alterations of the reflectance characteristics of everolimus-eluting bioresorbable vascular scaffold (BVS) struts have been reported in humans. In the absence of histology, the interpretation of the appearances of the struts by OCT remains speculative. We therefore report OCT findings with corresponding histology in the porcine coronary artery model immediately after and at 28 days and 2, 3, and 4 years after BVS implantation. METHODS AND RESULTS: Thirty-five polymeric BVS (3.0×12.0 mm) were singly implanted in the main coronary arteries of 17 pigs that underwent OCT and were then euthanized immediately (n=2), at 28 days (n=2), at 2 years (n=3), at 3 years (n=5), or at 4 years (n=5) after implantation. All BVS-implanted arteries in these animals were evaluated by histology except for 5 arteries examined at 2 years with gel permeation chromatography to assess the biodegradation of the polymeric device. Fourteen arteries with BVS from an additional 6 pigs were examined by gel permeation chromatography at 1 (n=1), 1.5 (n=2), and 3 (n=2) years. Corresponding OCT and histology images were selected with the distal and proximal radiopaque markers used as landmarks. At 28 days, by OCT, 82% of struts showed sharply defined, bright reflection borders, best described as a box-shaped appearance. Histologically, all struts appeared intact with no evidence of resorption. At 2 years, by OCT, 60±20 struts were discernible per BVS with 80.4% of the strut sites as a box-shaped appearance. Despite their defined appearance by OCT, by histology, these structures appeared to be composed of proteoglycan, with polymeric material being at such low level as to be no longer quantifiable by chromatography. At 3 years, by OCT, recognizable struts decreased to 28±9 struts per BVS: 43.7% showed dissolved black box; 34.8%, dissolved bright box; 16.1%, open box; and 5.4%, preserved box appearance. Histology shows that connective tissue cells within a proteoglycan-rich matrix replaced the areas previously occupied by the polymeric struts and coalesced into the arterial wall. At 4 years, by OCT, 10±6 struts were recognizable as either dissolved black or dissolved bright box. In histology, these struts are minimally discernible as foci of low-cellular-density connective tissue. Relative to the prediction of histological type by OCT appearance, the preserved box appearance of OCT corresponds well with 2-year histology (86.4%), whereas the dissolved bright and black box appearances correspond to 3-year histology (88.0% and 90.7%, respectively). Struts indiscernible by OCT correspond to the integrated strut footprints seen at 4 years (100%). CONCLUSIONS: Struts that are still discernible by OCT at 2 years are compatible with largely bioresorbed struts, as demonstrated by histological and gel permeation chromatography analysis. At 3 and 4 years, both OCT and histology confirm complete integration of the struts into the arterial wall.


Subject(s)
Absorbable Implants , Coronary Vessels/pathology , Drug-Eluting Stents , Sirolimus/analogs & derivatives , Tissue Scaffolds , Tomography, Optical Coherence/methods , Animals , Everolimus , Humans , Models, Animal , Polymers , Proteoglycans , Retrospective Studies , Swine , Swine, Miniature , Time Factors
5.
J Chromatogr Sci ; 46(1): 23-9, 2008 Jan.
Article in English | MEDLINE | ID: mdl-18218184

ABSTRACT

A novel sensitive high-throughput high-performance liquid chromatography assay is developed and validated for the simultaneous determination of everolimus and clobetasol propionate in pharmaceutical formulations. The chromatographic separation is achieved on a Zorbax Eclipse XDB-C18 reversed-phase column using a gradient elution, with solvent A: ammonium acetate (pH 6.8; 0.01 M) and solvent B: acetonitrile. The mean recovery ranges from 95.1% to 100.0% for clobetasol propionate and from 97.9% to 103.7% for everolimus. The limit of quantitation for each analyte is 0.02 microg/mL. The percent relative standard deviations are less than 3% for intra- and inter-day analyses. The proposed method can be used for the routine quality control of everolimus and clobetasol propionate in complex pharmaceutical formulations, especially the drug-delivery systems with a low total drug-load.


Subject(s)
Chromatography, High Pressure Liquid/methods , Clobetasol/analysis , Sirolimus/analogs & derivatives , Clobetasol/chemistry , Everolimus , Molecular Structure , Reproducibility of Results , Sirolimus/analysis , Sirolimus/chemistry
6.
J Biomed Mater Res B Appl Biomater ; 106(5): 1721-1729, 2018 07.
Article in English | MEDLINE | ID: mdl-28884951

ABSTRACT

The polymers poly(vinylidene fluoride-co-hexafluoropropylene) (PVDF-HFP) and poly(n-butyl methacrylate) (PBMA) are employed in manufacturing the XIENCE family of coronary stents. PBMA serves as a primer and adheres to both the stent and the drug coating. PVDF-HFP is employed in the drug matrix layer to hold the drug everolimus on the stent and control its release. Chemical stability of the polymers of XIENCE stents in the in-vivo environment was evaluated by pyrolysis-gas chromatography with mass spectrometry (Py-GC/MS) detection. For this evaluation, XIENCE stents explanted from porcine coronary arteries and from human coronary artery specimens at autopsy after 2-4 and 5-7 years of implantation, respectively, were compared to freshly manufactured XIENCE stents (controls). The comparison of pyrograms of explanted stent samples and controls showed identical fragmentation fingerprints of polymers, indicating that PVDF-HFP and PBMA maintained their chemical integrity after multiple years of XIENCE coronary stent implantation. The findings of the present study demonstrate the chemical stability of PVDF-HFP and PBMA polymers of the XIENCE family of coronary stents in the in-vivo environment, and constitute a further proof of the suitability of PVDF-HFP as a drug carrier for the drug eluting stent applications. © 2017 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 106B: 1721-1729, 2018.


Subject(s)
Coronary Vessels , Drug-Eluting Stents , Everolimus , Materials Testing , Animals , Coronary Vessels/metabolism , Coronary Vessels/pathology , Coronary Vessels/surgery , Everolimus/chemistry , Everolimus/pharmacokinetics , Everolimus/pharmacology , Female , Humans , Male , Swine
7.
Int J Cardiovasc Imaging ; 31(3): 471-82, 2015 Mar.
Article in English | MEDLINE | ID: mdl-25627777

ABSTRACT

The objective of the study is to validate intravascular quantitative echogenicity as a surrogate for molecular weight assessment of poly-l-lactide-acid (PLLA) bioresorbable scaffold (Absorb BVS, Abbott Vascular, Santa Clara, California). We analyzed at 9 time points (from 1- to 42-month follow-up) a population of 40 pigs that received 97 Absorb scaffolds. The treated regions were analyzed by echogenicity using adventitia as reference, and were categorized as more (hyperechogenic or upperechogenic) or less bright (hypoechogenic) than the reference. The volumes of echogenicity categories were correlated with the measurements of molecular weight (Mw) by gel permeation chromatography. Scaffold struts appeared as high echogenic structures. The quantification of grey level intensity in the scaffold-vessel compartment had strong correlation with the scaffold Mw: hyperechogenicity (correlation coefficient = 0.75; P < 0.01), upperechogenicity (correlation coefficient = 0.63; P < 0.01) and hyper + upperechogenicity (correlation coefficient = 0.78; P < 0.01). In the linear regression, the R(2) for high echogenicity and Mw was 0.57 for the combination of hyper and upper echogenicity. IVUS high intensity grey level quantification is correlated to Absorb BVS residual molecular weight and can be used as a surrogate for the monitoring of the degradation of semi-crystalline polymers scaffolds.


Subject(s)
Absorbable Implants , Angioplasty, Balloon, Coronary/instrumentation , Cardiovascular Agents/administration & dosage , Coronary Vessels/diagnostic imaging , Drug-Eluting Stents , Everolimus/administration & dosage , Ultrasonography, Interventional , Animals , Crystallization , Lactic Acid/chemistry , Linear Models , Models, Animal , Observer Variation , Polyesters , Polymers/chemistry , Predictive Value of Tests , Prosthesis Design , Reproducibility of Results , Swine , Time Factors
8.
Article in English | MEDLINE | ID: mdl-15358319

ABSTRACT

There are few methods available for the rapid and precise quantitation of non-covalent aggregation. Size-exclusion chromatography (SEC), a traditional approach, used to measure the non-covalent aggregation can easily disrupt the weak forces holding an aggregate together. Under the conditions described in this paper the disaggregation of non-covalent aggregate of the synthetic human parathyroid hormone hPTH (1-34) due to hydrophobic/electrostatic interactions with the size-exclusion chromatography column packing was completely suppressed. This report details the effectiveness of adding salts and organic solvents in the mobile phase to overcome non-specific interactions that disrupt the aggregate during the SEC process and may aid in the understanding precise quantitation of non-covalent aggregation.


Subject(s)
Teriparatide/analysis , Amino Acid Sequence , Chromatography, Gel , Humans , Indicators and Reagents , Light , Molecular Sequence Data , Scattering, Radiation , Sodium Chloride , Solvents , Spectrophotometry, Ultraviolet
9.
J Pharm Biomed Anal ; 34(1): 123-8, 2004 Jan 27.
Article in English | MEDLINE | ID: mdl-14738926

ABSTRACT

A high-performance liquid chromatography (HPLC) method has been developed and validated for the simultaneous determination of n-propionyl-p-aminophenol, 3-chloro-4-hydroxyacetanilide, 4'-hydroxyacetophenone, 4-hydroxyacetophenone oxime, 4-acetoxyacetanilide and 4'-chloroacetanilide, the main impurities in acetaminophen drug substance. The chromatographic separation was achieved on an Eclipse XDB-18 reversed-phase column using a gradient elution, being solvent A: 0.01 M phosphate buffer at pH 3.0 and solvent B: methanol. The limit of quantitation (S/N=10:1) was 0.1 microg/ml for each impurity. The coefficients of variation were less than 4% for intra-day and inter-day analyses. The individual recovery of acetaminophen spiked samples ranged from 94 to 104% and the mean recovery for each level from 99 to 103% in the 1-150 microg/ml range for all impurities. The proposed method was successfully applied to the analyses of different lots and different manufactures of acetaminophen drug substance. The proposed method can be used for the routine quality control of acetaminophen.


Subject(s)
Acetaminophen/analysis , Drug Contamination , Acetaminophen/chemistry , Chromatography, High Pressure Liquid/methods
10.
Circ Cardiovasc Interv ; 7(3): 330-42, 2014 Jun.
Article in English | MEDLINE | ID: mdl-24895447

ABSTRACT

BACKGROUND: The Absorb everolimus-eluting bioresorbable vascular scaffold (Absorb) has shown promising clinical results; however, only limited preclinical data have been published. We sought to investigate detailed pathological responses to the Absorb versus XIENCE V (XV) in a porcine coronary model with duration of implant extending from 1 to 42 months. METHODS AND RESULTS: A total of 335 devices (263 Absorb and 72 XV) were implanted in 2 or 3 main coronary arteries of 136 nonatherosclerotic swine and examined by light microscopy, scanning electron microscopy, pharmacokinetics, and gel permeation chromatography analyses at various time points. Vascular responses to Absorb and XV were largely comparable at all time points, with struts being sequestered within the neointima. Inflammation was mild to moderate (with absence of inflammation at 1 month) for both devices, although the scores were greater in Absorb at 6 to 36 months. Percent area stenosis was significantly greater in Absorb than XV at all time points except at 3 months. The extent of fibrin deposition was similar between Absorb and XV, which peaked at 1 month and decreased rapidly thereafter. Histomorphometry showed expansile remodeling of Absorb-implanted arteries starting after 12 months, and lumen area was significantly greater in Absorb than XV at 36 and 42 months. These changes correlated with dismantling of Absorb seen after 12 months. Gel permeation chromatography analysis confirmed that degradation of Absorb was complete by 36 months. CONCLUSIONS: Absorb demonstrates comparable long-term safety to XV in porcine coronary arteries with mild to moderate inflammation. Although Absorb was associated with greater percent stenosis relative to XV, expansile remodeling was observed after 12 months in Absorb with significantly greater lumen area at ≥ 36 months. Resorption is considered complete at 36 months.


Subject(s)
Absorbable Implants/adverse effects , Chromium Alloys/adverse effects , Coronary Vessels/pathology , Drug-Eluting Stents/adverse effects , Sirolimus/analogs & derivatives , Stents/adverse effects , Tissue Scaffolds/adverse effects , Animals , Coronary Angiography , Coronary Vessels/diagnostic imaging , Coronary Vessels/ultrastructure , Everolimus , Incidence , Microscopy, Electron, Scanning , Models, Animal , Neointima/diagnostic imaging , Neointima/pathology , Sirolimus/adverse effects , Sirolimus/pharmacokinetics , Swine , Swine, Miniature , Time Factors , Vasculitis/epidemiology
11.
J Pharm Biomed Anal ; 70: 94-100, 2012 Nov.
Article in English | MEDLINE | ID: mdl-22726455

ABSTRACT

High-throughput 96-well solid phase extraction (SPE) plate with C-18 reversed phase sorbent followed by UV-visible (UV-Vis) microplate reader was applied to the analysis of hydrophobic drugs in surfactant-containing dissolution media, which are often used to evaluate the in-vitro drug release of drug eluting stents (DES). Everolimus and dissolution medium containing Triton X-405 were selected as representatives, and the appropriate SPE conditions (adsorption, washing and elution) were investigated to obtain a practical and reliable sample clean-up. It was shown that the developed SPE procedure was capable of removing interfering components (Triton X-405 and its impurities), allowing for an accurate automated spectrophotometric analysis to be performed. The proposed UV-Vis spectrophotometric method yielded equivalent results compared to a classical LC analysis method. Linear regression analysis indicated that both methods have the ability to obtain test results that are directly proportional to the concentration of analyte in the sample within the selected range of 1.0-10 µg/ml for everolimus, with a coefficient of correlation (r(2)) value of >0.998 and standard deviation of the residuals (Syx) of <2%. The individual recoveries of everolimus ranged from 97 to 104% for the UV-Vis spectrophotometric method and from 98 to 102 for the HPLC method, respectively. The 95% CI of the mean recovery for the UV-Vis spectrophotometric method was 99-102% and for the HPLC method was 99-101%. No statistical difference was found between the mean recoveries of the methods (p=0.42). Hence the methods are free from interference due to Triton and other chemicals present in the dissolution medium. The variation in the amount of everolimus estimated by UV-Vis spectrophotometric and HPLC methods was ≤3.5%, and the drug release profiles obtained by both methods were found to be equivalent by evaluation with two-one-sided t-test (two-tailed, p=0.62; mean of differences, 0.17; 95% CI, 0.62-0.96) and similarity factor f2 (f2 value, 87). The excellent conformity of the results makes UV-Vis spectrophotometer an ideal tool for analyzing the drugs in the media containing surfactants, after SPE. The 96-well SPE plates in combination with UV-Vis microplate reader provide a high throughput method for the determination of in-vitro drug release profile of DES. Switching from HPLC to UV-Vis spectrophotometer microplate reader assay reduces the solvent consumption and labor required for the sample analyses. This directly impacts the profitability of the laboratory.


Subject(s)
Cardiovascular Agents/analysis , Polyethylene Glycols/chemistry , Sirolimus/analogs & derivatives , Solid Phase Extraction , Spectrophotometry, Ultraviolet , Surface-Active Agents/chemistry , Calibration , Cardiovascular Agents/chemistry , Chromatography, High Pressure Liquid , Drug Stability , Equipment Design , Everolimus , High-Throughput Screening Assays , Hydrophobic and Hydrophilic Interactions , Kinetics , Limit of Detection , Linear Models , Reference Standards , Reproducibility of Results , Sirolimus/analysis , Sirolimus/chemistry , Solid Phase Extraction/instrumentation , Solid Phase Extraction/standards , Solubility , Spectrophotometry, Ultraviolet/instrumentation , Spectrophotometry, Ultraviolet/standards
12.
Eur J Pharm Sci ; 42(3): 230-7, 2011 Feb 14.
Article in English | MEDLINE | ID: mdl-21126583

ABSTRACT

Accuracy and reliability of the analytical results are crucial for ensuring quality, safety and efficacy of drug eluting stents (DESs). Method validation is the process used to confirm that the analytical procedure employed for a specific test is suitable for its intended use. Results from method validation can be used to judge the quality, reliability and consistency of analytical results. Validation of analytical methods includes the identification of the performance parameters relevant for the given procedure, the definition of appropriate acceptance criteria and the appropriate design of the validation studies. Achieving an appropriate consideration of the analytical variability in assay procedures and setting acceptance criteria for analytical validations is however much more difficult than usually described. Criteria which are too wide may lead to unnecessary and incorrect out-of-specification (OOS) cases, resulting in bad reject decision for products. This study concentrates on analysis, through simulation, of the relation of method variability with specification limits for the total loaded dose of the active substance on the DES. The findings of this study point what levels of precision and accuracy are needed, in other words what is the magnitude of the allowable total error from all possible effects (both systematic and random) in an assay method in order to achieve the level of performance required for the methods applied routinely for the evaluation of the total loaded dose of DES as part of lot release/stability testing.


Subject(s)
Drug Administration Routes , Stents , Monte Carlo Method , Reproducibility of Results
13.
J Chromatogr B Analyt Technol Biomed Life Sci ; 877(20-21): 1946-50, 2009 Jul 01.
Article in English | MEDLINE | ID: mdl-19481508

ABSTRACT

The apolipoprotein A-I mimetic peptide D-4F is a potential therapeutical agent effective in maintaining cardiovascular health. A bioanalytical assay based on high-performance liquid chromatography coupled with electrospray ionization tandem mass spectrometry (LC/MS/MS) to quantitate the D-4F amount in rabbit plasma was developed and validated. A compound with a close structure similarity to the D-4F (only one amino acid A-V altered) was used as an internal standard. Both D-4F and the internal standard were extracted by protein precipitation using acetonitrile/0.2% Triton XL 80N. The correlation coefficient of the calibration curve was 0.9991 in the range 20-40,000 ng/mL. This assay can be used for pharmacokinetic studies of the drug. Also, it may be adjusted for the quantification of other members of apolipoprotein A-I mimetic peptide family.


Subject(s)
Anticholesteremic Agents/blood , Apolipoprotein A-I/blood , Chromatography, High Pressure Liquid/methods , Spectrometry, Mass, Electrospray Ionization/methods , Animals , Anticholesteremic Agents/pharmacokinetics , Apolipoprotein A-I/pharmacokinetics , Rabbits , Sensitivity and Specificity
14.
Eur J Pharm Sci ; 37(3-4): 217-22, 2009 Jun 28.
Article in English | MEDLINE | ID: mdl-19491008

ABSTRACT

The major objective of the present study was to develop an accelerated in vitro release method for everolimus/poly(lactic-co-glycolic acid) (PLGA) biodegradable DES that reflects and discriminates between many different sources of variations in the manufacturing process by introducing organic solvents in the release medium. To get further insight into the underlying drug release mechanisms, alongside release studies, the surface changes of the coated stents and the molecular weight changes of the polymer upon immersion in the selected release media were examined by scanning electron microscopy and size exclusion chromatography. The incorporation of acetonitrile in the release medium resulted in an increase in the drug release rate due to an increment in total porosity of the matrices. The developed method reflected and discriminated between different sources of variations in the manufacturing process and correlated with the real-time release. Over 80% of everolimus release occurred within 24h. The molecular and gravimetric weights of PLGA remained unchanged throughout the dissolution period, suggesting that the polymer does not undergo degradation through cleavage of its backbone ester linkages. It is likely that the drug release occurred mainly through its diffusion. The method can be employed as a rapid quality control test during development or commercial manufacturing.


Subject(s)
Absorbable Implants , Drug-Eluting Stents , Pharmaceutical Preparations/administration & dosage , Chromatography, Gel , Electrons , Everolimus , Excipients , Immunosuppressive Agents/administration & dosage , Immunosuppressive Agents/analysis , Immunosuppressive Agents/radiation effects , Lactic Acid , Microscopy, Electron, Scanning , Molecular Weight , Pharmaceutical Preparations/analysis , Polyglycolic Acid , Polylactic Acid-Polyglycolic Acid Copolymer , Sirolimus/administration & dosage , Sirolimus/analogs & derivatives , Sirolimus/analysis , Sirolimus/radiation effects , Solubility , Solvents
15.
J Infect Dis ; 187(4): 705-9, 2003 Feb 15.
Article in English | MEDLINE | ID: mdl-12599092

ABSTRACT

Production of the proinflammatory cytokines interleukin (IL)-1 alpha and tumor necrosis factor (TNF)-alpha and of the chemotactic chemokine macrophage inflammatory protein (MIP)-1 alpha by bronchoalveolar macrophages (BAMs) from mice in response to Aspergillus conidia was tested after in vivo administration of saline, dexamethasone, cortisone acetate, granulocyte-macrophage colony-stimulating factor (GM-CSF), or a combination. Dexamethasone suppressed production of IL-1 alpha, TNF-alpha, and MIP-1 alpha; GM-CSF reduced secretion slightly but antagonized dexamethasone suppression when the two were given in combination. Cortisone acetate gave results similar to dexamethasone, but cortisone acetate suppression of BAM responses lasted 7 days, > or = 4 days longer than dexamethasone suppression. The effect of GM-CSF on cortisone acetate suppression lasted at least 7 days. GM-CSF could promote resistance to conidia by maintaining proinflammatory responses.


Subject(s)
Aspergillosis/drug therapy , Aspergillosis/immunology , Bronchoalveolar Lavage Fluid/immunology , Cortisone/analogs & derivatives , Dexamethasone/pharmacology , Granulocyte-Macrophage Colony-Stimulating Factor/pharmacology , Interleukin-1/biosynthesis , Macrophages, Alveolar/immunology , Tumor Necrosis Factor-alpha/biosynthesis , Animals , Cells, Cultured , Chemokine CCL4 , Cortisone/pharmacology , Disease Models, Animal , Down-Regulation , Drug Therapy, Combination , Macrophage Inflammatory Proteins/biosynthesis , Macrophages, Alveolar/drug effects , Mice , Time Factors
16.
Cytokine ; 19(1): 14-20, 2002 Jul 07.
Article in English | MEDLINE | ID: mdl-12200108

ABSTRACT

There is substantial evidence that local production of proinflammatory cytokines are very important in host resistance to aspergillosis. Dexamethasone (DEX) down-regulates production of these cytokines by stimulated bronchoalveolar macrophages (BAM) and constitutes a risk factor for aspergillosis. Granulocyte-macrophage colony-stimulating factor (GM-CSF) antagonizes DEX suppression of antifungal activity by BAM. Here we investigated the possibility that GM-CSF could antagonize DEX down-regulation of interleukin (IL)-1alpha and tumour necrosis factor (TNF)-alpha production by stimulated BAM. Control BAM responded to increasing numbers of conidia of Aspergillus fumigatus with increasing production of IL-1 and TNF. DEX (10(-7)M) significantly suppressed IL-1 and TNF production by BAM+conidia. Although GM-CSF did not enhance IL-1 or TNF production by BAM+conidia, GM-CSF significantly antagonized DEX suppression of IL-1 cytokine production. For comparative purposes, lipopolysaccharide (LPS, 1 microg/ml) was used to stimulate BAM in experiments similar to the above. In contrast to the findings with conidia, GM-CSF enhanced the production of IL-1 (5-fold) and TNF (1.5-fold) by LPS treated BAM. DEX suppression of cytokine production by BAM+LPS was modestly but significantly antagonized by GM-CSF. Moreover, differences between regulation of IL-1 and TNF production by BAM+conidia or LPS and peritoneal macrophages (PM)+conidia or LPS were documented. Finally, the anti-inflammatory cytokine IL-10 was minimally produced by BAM + conidia or LPS, but IL-10 was produced by PM + conidia or LPS. In summary, these data indicate that the risk factor for aspergillosis associated with DEX could be lessened in the pulmonary compartment with GM-CSF. On the other hand, desired effects of DEX could be maintained in other compartments.


Subject(s)
Aspergillus/metabolism , Cytokines/biosynthesis , Dexamethasone/pharmacology , Granulocyte-Macrophage Colony-Stimulating Factor/pharmacology , Lipopolysaccharides/metabolism , Lung/cytology , Macrophages/metabolism , Animals , Enzyme-Linked Immunosorbent Assay , Glucocorticoids/pharmacology , Humans , Interleukin-1/metabolism , Interleukin-10/biosynthesis , Male , Mice , Tumor Necrosis Factor-alpha/metabolism
SELECTION OF CITATIONS
SEARCH DETAIL