Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters

Database
Language
Affiliation country
Publication year range
1.
Med Phys ; 34(3): 952-8, 2007 Mar.
Article in English | MEDLINE | ID: mdl-17441241

ABSTRACT

Very high dose per pulse (3-13 cGy/pulse) high energy electron beams are currently produced by special linear accelerators (linac) dedicated to Intra Operative Radiation Therapy (IORT). The electron beams produced by such linacs are collimated by special Perspex applicators of various size and cylindrically shaped. The biggest problems from the dosimetric point of view are caused by the high dose-per-pulse values and the use of inclined applicators. In this work measurements of absolute dose for the inclined applicators were done by using a small cylindrical ionization chamber, type CC01 (Wellhofer), a parallel plane ionization chamber type Markus (PTW 23343) and radiochromic films type EBT. We show a method which allows calculating the quality correction factors for CC01 chamber with an uncertainty of 1% and the absolute dose value for the inclined applicators using CC01 with an uncertainty of 3.1% for electron beams of energy of 6 and 7 MeV produced by the linac dedicated to IORT Novac7.


Subject(s)
Film Dosimetry/methods , Radiometry/methods , Radiotherapy, High-Energy/methods , Radiotherapy/methods , X-Ray Film , Calibration , Electrons , Humans , Ions , Kinetics , Models, Statistical , Particle Accelerators
SELECTION OF CITATIONS
SEARCH DETAIL