1.
J Cardiovasc Electrophysiol ; 31(5): 1213-1217, 2020 05.
Article in English | MEDLINE | ID: mdl-32108404

ABSTRACT

A 25-year-old runner received a single-lead, VDD pacemaker after ablation of AV nodal reentrant tachycardia complicated by intermittent AV block. The rate-adaptive AV delay algorithm (RAAV), which shortens the sensed AV interval (SAV) at faster atrial rates, was programmed to provide a physiologic SAV with exercise. She developed repetitive, atypical, long-RP pacemaker-mediated tachycardia (PMT) because the RAAV shortened the antegrade SAV and retrograde conduction occurred over the slow AV nodal pathway. PMT was refractory to usual programming solutions. Using high-density electroanatomic mapping, we were able to ablate the retrograde limb of PMT without further damaging AV conduction.


Subject(s)
Atrioventricular Block/therapy , Cardiac Pacing, Artificial/adverse effects , Catheter Ablation/adverse effects , Heart Conduction System/surgery , Pacemaker, Artificial , Tachycardia, Ventricular/surgery , Action Potentials , Adult , Atrioventricular Block/diagnosis , Atrioventricular Block/etiology , Atrioventricular Block/physiopathology , Electrocardiography, Ambulatory , Electrophysiologic Techniques, Cardiac , Female , Heart Conduction System/physiopathology , Heart Rate , Humans , Reoperation , Running , Tachycardia, Ventricular/diagnosis , Tachycardia, Ventricular/physiopathology , Time Factors , Treatment Outcome
2.
Pacing Clin Electrophysiol ; 42(4): 478-482, 2019 Apr.
Article in English | MEDLINE | ID: mdl-30515880

ABSTRACT

SecureSense is an implantable cardioverter defibrillator algorithm that differentiates lead-related oversensing from ventricular tachycardia/ventricular fibrillation by continuous comparison between the near-field (NF) and the far-field (FF) electrogram. If lead noise is identified, inappropriate therapy is withheld. Undersensing on the FF channel could result in inappropriate inhibition of life-saving therapy. Thus, the device automatically switches SecureSense to passive mode if undersensing on the FF channel is suspected. We report here the first cases of inappropriate automatic SecureSense deactivation due to misdiagnosed FF undersensing in pacemaker-dependent patients. Physicians should be aware that SecureSense does not withhold an inappropriate therapy for sustained oversensing in pacemaker-dependent patients.


Subject(s)
Algorithms , Defibrillators, Implantable , Equipment Failure Analysis , Tachycardia, Ventricular/therapy , Ventricular Fibrillation/therapy , Aged , Aged, 80 and over , Electrocardiography , Equipment Design , Humans , Male , Tachycardia, Ventricular/diagnosis , Ventricular Fibrillation/diagnosis
3.
J Cardiovasc Electrophysiol ; 29(8): 1125-1134, 2018 08.
Article in English | MEDLINE | ID: mdl-29858871

ABSTRACT

INTRODUCTION: Recognition of implantable cardioverter defibrillator (ICD) lead malfunction before occurrence of life threatening complications is crucial. We aimed to assess the effectiveness of remote monitoring associated or not with a lead noise alert for early detection of ICD lead failure. METHODS: From October 2013 to April 2017, a median of 1,224 (578-1,958) ICD patients were remotely monitored with comprehensive analysis of all transmitted materials. ICD lead failure and subsequent device interventions were prospectively collected in patients with (RMLN) and without (RM) a lead noise alert (Abbott Secure Sense™ or Medtronic Lead Integrity Alert™) in their remote monitoring system. RESULTS: During a follow-up of 4,457 patient years, 64 lead failures were diagnosed. Sixty-one (95%) of the diagnoses were made before any clinical complication occurred. Inappropriate shocks were delivered in only one patient of each group (3%), with an annual rate of 0.04%. All high voltage conductor failures were identified remotely by a dedicated impedance alert in 10 patients. Pace-sense component failures were correctly identified by a dedicated alert in 77% (17 of 22) of the RMLN group versus 25% (8 of 32) of the RM group (P = 0.002). The absence of a lead noise alert was associated with a 16-fold increase in the likelihood of initiating either a shock or ATP (OR: 16.0, 95% CI 1.8-143.3; P = 0.01). CONCLUSION: ICD remote monitoring with systematic review of all transmitted data is associated with a very low rate of inappropriate shocks related to lead failure. Dedicated noise alerts further reduce inappropriate detection of ventricular arrhythmias.


Subject(s)
Clinical Alarms , Defibrillators, Implantable/adverse effects , Electrodes, Implanted/adverse effects , Equipment Failure Analysis/methods , Equipment Failure , Remote Sensing Technology/methods , Adult , Aged , Clinical Alarms/trends , Defibrillators, Implantable/trends , Electrodes, Implanted/trends , Female , Follow-Up Studies , Humans , Male , Middle Aged , Prospective Studies
4.
J Cardiovasc Electrophysiol ; 28(2): 224-232, 2017 02.
Article in English | MEDLINE | ID: mdl-27957764

ABSTRACT

BACKGROUND: Implantable cardioverter defibrillators (ICDs) must establish a balance between delivering appropriate shocks for ventricular tachyarrhythmias and withholding inappropriate shocks for lead-related oversensing ("noise"). To improve the specificity of ICD therapy, manufacturers have developed proprietary algorithms that detect lead noise. The SecureSenseTM RV Lead Noise discrimination (St. Jude Medical, St. Paul, MN, USA) algorithm is designed to differentiate oversensing due to lead failure from ventricular tachyarrhythmias and withhold therapies in the presence of sustained lead-related oversensing. METHODS AND RESULTS: We report 5 patients in whom appropriate ICD therapy was withheld due to the operation of the SecureSense algorithm and explain the mechanism for inhibition of therapy in each case. Limitations of algorithms designed to increase ICD therapy specificity, especially for the SecureSense algorithm, are analyzed. CONCLUSION: The SecureSense algorithm can withhold appropriate therapies for ventricular arrhythmias due to design and programming limitations. Electrophysiologists should have a thorough understanding of the SecureSense algorithm before routinely programming it and understand the implications for ventricular arrhythmia misclassification.


Subject(s)
Algorithms , Arrhythmias, Cardiac/diagnosis , Arrhythmias, Cardiac/therapy , Electrophysiologic Techniques, Cardiac/methods , Signal Processing, Computer-Assisted , Signal-To-Noise Ratio , Aged , Arrhythmias, Cardiac/physiopathology , Defibrillators, Implantable , Electric Countershock , Equipment Failure , Female , Humans , Predictive Value of Tests , Prosthesis Design , Reproducibility of Results , Software Design , Treatment Outcome
5.
Pacing Clin Electrophysiol ; 39(7): 774-778, 2016 Jul.
Article in English | MEDLINE | ID: mdl-27378562

ABSTRACT

Diaphragmatic myopotential oversensing (DMO) causes inhibition of pacing and inappropriate detection of ventricular fibrillation in implantable cardioverter defibrillators (ICDs). It occurs almost exclusively with integrated bipolar leads and is extremely rare with dedicated bipolar leads. If DMO cannot be corrected by reducing programmed sensitivity, ventricular lead revision is often required. The new Low Frequency Attenuation (LFA) filter in St. Jude Medical ICDs (St. Jude Medical, Sylmar, CA, USA) alters the sensing bandpass to reduce T-wave oversensing. This paper aims to present the LFA filter as a reversible cause of DMO. Unnecessary lead revision can be avoided by the simple programming solution of deactivating this LFA filter.

7.
JACC Clin Electrophysiol ; 10(3): 604-610, 2024 Mar.
Article in English | MEDLINE | ID: mdl-38276928

ABSTRACT

Recently, polarity-dependent shock failures were reported in implantable cardioverter-defibrillators caused by structural failure in the high-voltage feedthrough. Short circuits may occur when the right ventricular coil is cathodal for phase 1 of biphasic shocks (cathodal shock). This viewpoint proposes a mechanism for observed polarity dependence and considers whether the same mechanism may apply in other shock-induced, short circuits. Implantable cardioverter-defibrillator connections to the lead traverse feedthroughs into the hermetically sealed housing ("Can"). The feedthrough comprises 2 concentric, conducting metal cylinders, the inner pin-conductor to the right ventricular coil and outer Can, separated by impermeable insulation. Shock failure depends on 3 conditions: 1) development of a fluid layer in the feedthrough, creating a conduction path in parallel with the shock pathway; 2) the radial gradient of the electric field in the fluid, so resistive heating during a shock vaporizes water to form a high-resistance gas bubble around the pin; and 3) field emission of electrons at the cathode, with rate and energy dependent on the field's strength and the cathode's potential-energy barrier to emission. For cathodal shocks, electrons emitted at the metal pin may initiate an ionization avalanche in the gas until it "breaks down" into a low-resistance plasma, resulting in a short circuit. For anodal shocks, the effective cathode is the liquid-gas interface, where the field is weaker than at the pin. Additionally, solvated electrons in aqueous solution must overcome a higher potential-energy barrier to be emitted. This permits the high-resistance gas bubble to stabilize so that the shock is completed.


Subject(s)
Defibrillators, Implantable , Humans , Defibrillators, Implantable/adverse effects , Heart Rate
8.
Article in English | MEDLINE | ID: mdl-38752961

ABSTRACT

BACKGROUND: The extravascular (EV) implantable cardioverter-defibrillator (ICD) includes features to address sensing and arrhythmia detection challenges presented by its substernal lead location. OBJECTIVES: In this study, the authors sought to evaluate sensing and detection performance in 299 patients discharged with an EV-ICD in the global pivotal study. METHODS: We reviewed and adjudicated all induced ventricular fibrillation (VF) episodes and spontaneous device-stored episodes that satisfied rate and duration criteria in a programmed ventricular tachycardia (VT)/VF therapy zone. RESULTS: At implantation, all EV-ICDs detected induced VF at the programmed sensitivity; 95.9% detected VF with a 3× safety margin. In follow-up, EV-ICDs detected all 59 VT/VF episodes that sustained until therapy. Of 1,034 non-VT/VF episodes, oversensing caused 87.9% and supraventricular tachycardia caused 12.1%. Therapy was withheld in 80.9%, aborted in 10.6%, and delivered in 8.5%. The most common causes of oversensing were myopotentials (61.2%) and P-wave oversensing (PWOS) (19.9%). Inappropriate shocks occurred in only 3.2% of myopotential episodes, but in 21.8% of PWOS episodes. Myopotential oversensing was more common with Ring-Can sensing (P < 0.0001) and correlated with low R-wave amplitude (P < 0.0001). PWOS occurred almost exclusively with Ring1-Ring2 sensing (P = 0.0001) and began with transient decrease in R-wave or increase in P-wave amplitude (P < 0.0001). In software emulation, a new PWOS discriminator significantly reduced total inappropriate detections. CONCLUSIONS: In a global population, EV-ICD detected induced and spontaneous VT/VF accurately. Although discriminators withheld detection from most non-VT/VF episodes, inappropriate shocks were common. The most common cause was PWOS, which may be reduced by optimizing sensing at implantation and incorporation of the PWOS discriminator, which is now in the current device. (Extravascular ICD Pivotal Study [EV ICD]; NCT04060680).

9.
Circulation ; 123(10): 1052-60, 2011 Mar 15.
Article in English | MEDLINE | ID: mdl-21357826

ABSTRACT

BACKGROUND: T-wave alternans (TWA) increases before ventricular tachycardia (VT) or fibrillation (VF), suggesting that it may warn of VT/VF in implantable cardioverter-defibrillator patients. Recently, we described a method for measuring alternans and nonalternans variability (TWA/V) from electrograms (EGMs) stored in implantable cardioverter-defibrillators before VT/VF. The goal of this prospective, multicenter study was to determine whether EGM TWA/V was greater before VT/VF than at baseline. METHODS AND RESULTS: We enrolled 63 implantable cardioverter-defibrillator patients. TWA/V was computed from stored EGMs before spontaneous VT/VF and from sequential windows of 8 pairs of beats using 4 different control recordings: baseline rhythm, rapid pacing at 105 bpm, segments of ambulatory Holter EGMs matched to the time of VT/VF episodes, and EGMs before spontaneous supraventricular tachycardia. During follow-up, 28 patients had 166 episodes of VT/VF. TWA/V was greater before VT/VF (62.9 ± 3.1 µV; n = 28) than during baseline rhythm (12.8 ± 1.8 µV; P < 0.0001; n = 62), during rapid pacing (14.5 ± 2.0 µV; P < 0.0001; n = 52), before supraventricular tachycardia (27.5 ± 6.1 µV; P < 0.0001; n = 9), or during time-matched ambulatory controls (12.3 ± 3.5 µV; P < 0.0001; n = 16). By logistic regression, the odds of VT/VF increased by a factor of 2.2 for each 10-µV increment in TWA/V (P < 0.0001). CONCLUSIONS: In implantable cardioverter-defibrillator patients, EGM TWA/V is greater before spontaneous VT/VF than in control recordings. Future implantable cardioverter-defibrillators that measure EGM TWA/V continuously may warn patients and initiate pacing therapies to prevent VT/VF.


Subject(s)
Defibrillators, Implantable , Electrophysiologic Techniques, Cardiac/methods , Tachycardia, Supraventricular/physiopathology , Tachycardia, Ventricular/physiopathology , Aged , Female , Humans , Male , Middle Aged , Prospective Studies , Tachycardia, Ventricular/diagnosis
10.
Heart Rhythm O2 ; 3(6Part A): 688-698, 2022 Dec.
Article in English | MEDLINE | ID: mdl-36589919

ABSTRACT

Background: Pace-sense conductors comprise a pacing coil to the tip electrode and cable to the ring-electrode. Implantable cardioverter-defibrillator (ICD) lead-monitoring diagnostics include pacing impedance (direct current resistance [DCR]) and measures of oversensing. How they change as fractures progress is unknown. Objectives: To characterize the relationship between oversensing, impedance, and structural changes in ICD leads developing pace-sense conductor fractures. Methods: We performed bending tests on 39 leads connected to ICD generators in an electrolyte bath with simulated electrograms. DCR was recorded every 3 minutes; electrograms were telemetered continuously. Twenty-two leads were tested to develop partial or complete fracture criteria confirmed by imaging, using DCR or DCR variability measured by standard deviation (σDCR). Results are reported for 17 other test leads. Results: Initial oversensing occurred with partial pacing coil fracture vs complete ring cable fracture and correlated with bending-induced DCR peaks. These peaks were too small to be detected by clinical impedance measurements and were characterized by small increases in σDCR (≥0.5 Ω). Impedance threshold alerts occurred at complete pacing coil fracture but only later for ring cable fractures. The oversensing alert triggered before device-detected ventricular fibrillation more frequently than impedance alerts (94% vs 17%; P = .00002). Conclusions: In conductor fracture, early oversensing corresponds to partial pacing coil fracture or complete ring cable fracture and correlates with transient bending-induced impedance increases, which are detected by impedance variability but too small to trigger clinical impedance alerts. This explains why clinical oversensing alerts provide more warning for device-detected ventricular fibrillation than impedance alerts and suggests how to improve impedance diagnostics based on short-term variability.

11.
Heart Rhythm ; 19(1): 154-164, 2022 01.
Article in English | MEDLINE | ID: mdl-34597770

ABSTRACT

Implantable cardioverter-defibrillators (ICDs) incorporate automated, lead-monitoring alerts (alerts) and other diagnostics to detect defibrillation lead failure (LF) and minimize its adverse clinical consequences. Partial conductor fractures cause oversensing, but pacing or high-voltage alerts for high impedance detect only complete conductor fracture. In both pacing and high-voltage insulation breaches, low-impedance alerts require complete breach with metal-to-metal contact. Oversensing alerts for pace-sense LF also require complete breach, but not metal-to metal contact. Electrograms (EGMs) from leads with confirmed fractures have characteristics findings. In insulation breach, however, oversensed EGMs reflect characteristics of the source signal. Oversensing alerts that operate on the sensing channel analyze R-R intervals for 2 patterns typical of LF but uncommon in other conditions: a rapidly increasing count of "nonphysiological" short intervals and rapid "nonsustained tachycardias." These alerts are sensitive but nonspecific. Alerts that compare sensing and shock channels define oversensing as sensed events that do not correlate temporally with EGMs on the shock channel. Their performance depends on implementation. Specific advantages and limitations are reviewed. Most ICDs measure impedance using subthreshold pulses. Patterns in impedance trends provide diagnostic information, whether or not an alert is triggered. Gradual increases in impedance do not indicate structural LF, but they may cause failed defibrillation if shock impedance is high enough. Because impedance-threshold alerts are insensitive, normal impedance trends never exclude LF, but an abrupt increase that triggers an alert almost always indicates a header connection issue or LF. Methods for discriminating connection issues from LF are reviewed.


Subject(s)
Defibrillators, Implantable , Equipment Failure Analysis , Electrocardiography , Humans
12.
Heart Rhythm ; 19(3): 363-371, 2022 03.
Article in English | MEDLINE | ID: mdl-34767985

ABSTRACT

BACKGROUND: Implantable cardioverter-defibrillator (ICD) lead monitoring diagnostic alerts facilitate the diagnosis of structural lead failure. OBJECTIVE: The purpose of this study was to prospectively study the performance of Medtronic ICD lead monitoring alerts. METHODS: A prespecified ancillary substudy, World-Wide Randomized Antibiotic Envelope Infection Prevention Trial, was conducted in patients with an ICD with all available alerts enabled. The investigators reported possible lead system events (LSEs), with or without an alert. An independent committee reviewed all data and classified events as lead failure, other LSE, or nonlead system events (NLEs). RESULTS: In 4942 patients who were followed for 19.4 ± 8.7 months, there were 124 alerts (65 LSEs, 59 NLEs) and 19 LSEs without an alert. Lead monitoring alerts had 100% sensitivity for the 48 adjudicated lead failures (95% confidence interval 92.6%-100%) and for 10 events adjudicated as either lead failure or connection issue. The positive predictive value of alerts for lead failure was 38.7% (48 of 124). For 34 pace-sense lead failures, an alert that incorporated oversensing was more sensitive than the pacing impedance threshold alert (33 patients [97.1%] vs 9 patients [26.5%]; P < .0001). However, the sensitivity was only 13.6% for lead dislodgments or perforations. Inappropriate shocks occurred in 2 patients with pace-sense lead failure (5.9%). No patient had unnecessary lead replacement for any of the NLEs. CONCLUSION: In this first real-world prospective study, lead monitoring alerts had 100% sensitivity for identifying lead failures. Although their positive predictive value was modest, no false-positive alerts resulted in an unnecessary lead replacement. For the diagnosis of pace-sense lead failure, an alert for oversensing was more sensitive than a pacing impedance threshold alert. TRIAL REGISTRATION: ClinicalTrials.gov identifier: NCT02277990.


Subject(s)
Defibrillators, Implantable , Defibrillators, Implantable/adverse effects , Electric Impedance , Equipment Failure , Humans , Predictive Value of Tests , Prospective Studies
13.
Circulation ; 121(4): 487-97, 2010 Feb 02.
Article in English | MEDLINE | ID: mdl-20083678

ABSTRACT

BACKGROUND: Inappropriate and unnecessary implantable cardioverter-defibrillator shocks continue to be highly prevalent. METHODS AND RESULTS: We prospectively evaluated a new algorithm for discriminating supraventricular (SVT) and ventricular (VT) tachycardias with 1:1 atrioventricular association that is based on the response of the arrhythmia to simultaneous or convergent dual-chamber antitachycardia pacing. Patients undergoing dual-chamber cardioverter-defibrillator implantation were randomized to the simultaneous atrioventricular and convergent atrioventricular arms with crossover at 3 months. Sixty-three patients had 1407 1:1 antitachycardia pacing sequences suitable for analysis (1381 1:1 SVT episodes in 32 patients and 26 1:1 VT episodes in 6 patients). Antitachycardia pacing terminated 66 of 1381 SVT (5%; generalized estimating equations adjusted, 23.8%) and 20 of 26 VT (77%; generalized estimating equation adjusted, 68.6%) episodes. After the exclusion of sinus tachycardia, the new software terminated 40 of 57 (70%; generalized estimating equation adjusted, 70.2%) SVT episodes. The new algorithm terminated or correctly classified 1379 of 1381 SVT sequences for an overall specificity of 99.9% (generalized estimating equation adjusted, 99.8%) and 23 of 26 VT for an overall sensitivity of 88.5% (generalized estimating equation adjusted, 82.1%). There were no statistically significant differences between the simultaneous and the convergent atrioventricular antitachycardia pacing sequences in their ability to confirm VT or reject SVT. No significant proarrhythmias were noted. CONCLUSIONS: We describe here a new pacing algorithm in dual-chamber defibrillators that can terminate arrhythmias or discriminate between 1:1 SVT and VT if the arrhythmia persists. Testing this new algorithm in larger patient populations is warranted. CLINICAL TRIAL REGISTRATION INFORMATION- URL: http://ftp.resource.org/gpo.gov/register/2007/2007_15297.pdf. IDE No. G060230.


Subject(s)
Algorithms , Arrhythmias, Cardiac/diagnosis , Defibrillators, Implantable , Tachycardia, Supraventricular/diagnosis , Tachycardia, Ventricular/diagnosis , Aged , Arrhythmias, Cardiac/physiopathology , Arrhythmias, Cardiac/therapy , Cross-Over Studies , Diagnosis, Differential , Electrocardiography , Female , Humans , Male , Middle Aged , Prospective Studies , Sensitivity and Specificity , Software , Tachycardia, Supraventricular/physiopathology , Tachycardia, Supraventricular/therapy , Tachycardia, Ventricular/physiopathology , Tachycardia, Ventricular/therapy
14.
Circulation ; 122(15): 1449-55, 2010 Oct 12.
Article in English | MEDLINE | ID: mdl-20876433

ABSTRACT

BACKGROUND: Downloadable software upgrades are common in consumer electronics but not in implantable medical devices. Fractures of implantable cardioverter-defibrillator (ICD) leads present commonly as inappropriate shocks. A lead-integrity alert (LIA) designed to reduce inappropriate shocks is the first software download approved to enhance nominally functioning, previously implanted ICDs. METHODS AND RESULTS: We performed a prospective study to determine whether an LIA could reduce inappropriate shocks. Patients were included if they had ICD lead fractures confirmed by analysis of explanted leads. The LIA group included the first 213 patients who met the inclusion criteria after the LIA was approved who had the LIA downloaded. The LIA is triggered either by high impedance or rapid oversensing. It responds by delaying detection of ventricular fibrillation and initiating a patient alert every 4 hours. The control group included the first 213 patients who did not have the LIA downloaded. They were monitored by conventional daily impedance measurements that respond with a daily alert. The LIA group had a 46% relative reduction (95% confidence interval 34% to 55%) in the percentage of patients with ≥1 inappropriate shock (LIA 38% versus control 70%, P<0.001) and a 50% relative reduction (95% confidence interval 33% to 61%) in the percentage with ≥5 shocks (25% versus 50%, P<0.001). The LIA group also had a higher percentage of patients who either did not receive a shock or had ≥3 days of warning before the shock (72% versus 50%, P<0.001). CONCLUSIONS: A software download that upgrades previously implanted ICDs without surgical revision reduces inappropriate shocks caused by lead fractures.


Subject(s)
Algorithms , Defibrillators, Implantable/adverse effects , Lead , Software , Ventricular Fibrillation/therapy , Adult , Aged , Case-Control Studies , Electrocardiography , Electrodes/adverse effects , Equipment Failure , Female , Humans , Male , Middle Aged , Predictive Value of Tests , Prospective Studies , Retrospective Studies
16.
JACC Clin Electrophysiol ; 7(11): 1387-1399, 2021 11.
Article in English | MEDLINE | ID: mdl-33933410

ABSTRACT

OBJECTIVES: This study reports the sensing and arrhythmia detection performance of a novel extravascular (EV) implantable cardioverter-defibrillator (ICD) in a first-in-human pilot study. BACKGROUND: The EV ICD lead is implanted in the substernal space, resulting in novel sensing and detection challenges. It uses a programmable sensing profile with new or modified discrimination of oversensing and of ventricular tachycardia (VT) from supraventricular tachycardia (SVT). METHODS: Electrograms were post-processed from induced ventricular fibrillation (VF) at implant to determine virtual detection times for each programmable sensitivity and the least-sensitive safe sensitivity setting. In ambulatory patients, programmed sensitivity provided at least a twofold safety margin for detecting induced VF. Noise discrimination was stress tested, and the effects of source, posture, and lead maturation were determined on electrogram amplitude. Telemetry Holter monitors were used to quantify undersensing and oversensing. RESULTS: In 20 patients at implant, the least-sensitive safe sensitivity for VF detection ranged from 0.1 to 0.6 mV. Seventeen patients were followed up for a total of 16.6 patient-years. Electrogram amplitudes were stable over time, but there were significant differences among postures and sensing vectors. For the primary sensing vector, the weighted oversensing and undersensing rates were 1.03% and 0.40% respectively, on a beat-to-beat basis. Oversensing did not cause inappropriate therapy in patients with in situ leads. Oversensing discriminators withheld VF detection in 4 of 5 spontaneous, sustained oversensed episodes. SVT-VT discriminators correctly classified 93% of 128 sustained SVTs in monitor zones. CONCLUSIONS: In the EV ICD pilot study, oversensing did not cause inappropriate therapy during ambulatory follow-up of stable leads.


Subject(s)
Defibrillators, Implantable , Tachycardia, Ventricular , Algorithms , Humans , Pilot Projects , Tachycardia, Ventricular/diagnosis , Tachycardia, Ventricular/therapy , Ventricular Fibrillation
17.
Circ Arrhythm Electrophysiol ; 13(2): e008092, 2020 02.
Article in English | MEDLINE | ID: mdl-31985260

ABSTRACT

Impedance is the ratio of voltage to current in an electrical circuit. Cardiovascular implantable electronic devices measure impedance to assess the structural integrity electrical performance of leads, typically using subthreshold pulses. We review determinants of impedance, how it is measured, variation in clinically measured pacing and high-voltage impedance and impedance trends as a diagnostic for lead failure and lead-device connection problems. We consider the differential diagnosis of abnormal impedance and the approach to the challenging problem of a single, abnormal impedance measurement. Present impedance provides a specific but insensitive diagnostic. For pacing circuits, we review the complementary roles of impedance and more sensitive oversensing diagnostics. Shock circuits lack a sensitive diagnostic. This deficiency is particularly important for insulation breaches, which may go undetected and present with short circuits during therapeutic shocks. We consider new methods for measuring impedance that may increase sensitivity for insulation breaches.


Subject(s)
Defibrillators, Implantable , Electric Impedance , Equipment Failure Analysis , Diagnosis, Differential , Humans
19.
Circulation ; 118(21): 2122-9, 2008 Nov 18.
Article in English | MEDLINE | ID: mdl-18981299

ABSTRACT

BACKGROUND: The primary method for monitoring implantable cardioverter-defibrillator lead integrity is periodic measurement of impedance. Sprint Fidelis leads are prone to pace-sense lead fractures, which commonly present as inappropriate shocks caused by oversensing. METHODS AND RESULTS: We developed and tested an algorithm to enhance early identification of lead fractures and to reduce inappropriate shocks. This lead-integrity algorithm, which can be downloaded into presently implanted implantable cardioverter-defibrillators, alerts the patient and/or physician when triggered by either oversensing or excessive increases in impedance. To reduce inappropriate shocks, the lead-integrity algorithm increases the number of intervals to detect (NID) ventricular fibrillation when triggered. The lead-integrity algorithm was tested on data from 15 970 patients with Fidelis leads (including 121 with clinically diagnosed fractures) and 95 other fractured leads confirmed by analysis of returned product. The effect of the NID on inappropriate shocks was tested in 92 patients with 927 shocks caused by lead fracture. Increasing the NID reduced inappropriate shocks (P<0.0001). The lead-integrity algorithm provided at least a 3-day warning of inappropriate shocks in 76% (95% CI, 66 to 84) of patients versus 55% (95% CI, 43 to 64) for optimal impedance monitoring (P=0.007). Its positive predictive value was 72% for lead fractures and 81% for lead fractures or header-connector problems requiring surgical intervention. The false-positive rate was 1 per 372 patient-years of monitoring. CONCLUSIONS: A lead-integrity algorithm developed for download into existing implantable cardioverter-defibrillators increases short-term warning of inappropriate shocks in patients with lead fractures and reduces the likelihood of inappropriate shocks. It is the first downloadable RAMware to enhance the performance of nominally functioning implantable cardioverter-defibrillators and the first implantable cardioverter-defibrillator monitoring feature that triggers real-time changes in ventricular fibrillation detection parameters to reduce inappropriate shocks.


Subject(s)
Algorithms , Defibrillators, Implantable , Monitoring, Physiologic/methods , Shock, Cardiogenic/prevention & control , Electric Impedance , Equipment Failure , Equipment Failure Analysis , Humans , Monitoring, Physiologic/instrumentation