Your browser doesn't support javascript.
loading
Microcalorimetric methods for substrate determination in flow systems with immobilized enzymes.
Biochim Biophys Acta ; 429(1): 283-90, 1976 Mar 11.
Article in En | MEDLINE | ID: mdl-1260032
The enthalpy of processes catalyzed by immobilized enzymes in the reaction cell of a LKB-flow calorimeter is used for determination of urea (0.5-5 mumol) and glucose (0.03-0.5 mumol). Accuracy is 2-5% and the time needed for one analysis is 20 min. A sensitive "enzyme thermistor" consisting of a flow through cell with an immobilized enzyme and two thermistors is described, which permits glucose determinations (0.05-1 mumol +/- 0.03 mumol) by means of temperature difference caused by reaction heat. Coupling of enzyme reactions for increasing reaction heat and consequently sensitivity in calorimetric determinations is demonstrated.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Calorimetry / Enzymes Language: En Journal: Biochim Biophys Acta Year: 1976 Type: Article