Stability of cerebral blood flow measures using a split-dose technique with (99m)Tc-exametazime SPECT.
Newberg, Andrew B; Saffer, Janet; Farrar, John; Pourdehnad, Michael; Alavi, Abass.
Affiliation
  • Newberg AB; Division of Nuclear Medicine, Department of Radiology, Hospital of the University of Pennsylvania, Philadelphia, PA 19104, USA. newberg@rad.upenn.edu
Nucl Med Commun ; 26(5): 475-8, 2005 May.
Article in En | MEDLINE | ID: mdl-15838432
ABSTRACT

AIM:

To determine whether there is stability of cerebral blood flow (CBF) measures using single photon emission computed tomography (SPECT) imaging in healthy controls in a test-retest split-dose paradigm. Such a paradigm is frequently used in the clinical and research setting to assess various brain states.

METHODS:

Five healthy volunteers underwent two brain SPECT scans after the administration of low and high doses of (99m)Tc-exametazime. The first SPECT scan was acquired approximately 30 min after the intravenous injection of approximately 259 MBq of (99m)Tc-exametazime. The second SPECT scan was acquired approximately 30 min after the intravenous injection of 925 MBq of (99m)Tc-exametazime. Both scans were acquired over approximately 30-45 min and the images were reconstructed using filtered backprojection, a low-pass filter and Chang's first-order attenuation correction. Values were obtained for regions of interest (ROIs) in major brain structures and normalized to whole brain activity. Counts on the second SPECT scan were also decay corrected for activity from the first scan.

RESULTS:

The results demonstrated a strong correlation between the low-dose and high-dose scans for all regions (r = 0.86, P<0.0001). Symmetries were preserved with a strong correlation between low-dose and high-dose scans (r = 0.70, P<0.0001). Finally, most regions demonstrated less than a 5% difference between the low-dose and high-dose scans.

CONCLUSIONS:

The results of this study demonstrate that the split-dose technique can be employed for clinical and research applications to measure CBF in different brain states using two SPECT scans on the same day.
Subject(s)
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Tomography, Emission-Computed, Single-Photon / Cerebrovascular Circulation / Technetium Tc 99m Exametazime Limits: Adult / Female / Humans / Male / Middle aged Language: En Journal: Nucl Med Commun Year: 2005 Type: Article Affiliation country: United States
Search on Google
Collection: 01-internacional Database: MEDLINE Main subject: Tomography, Emission-Computed, Single-Photon / Cerebrovascular Circulation / Technetium Tc 99m Exametazime Limits: Adult / Female / Humans / Male / Middle aged Language: En Journal: Nucl Med Commun Year: 2005 Type: Article Affiliation country: United States