The spontaneous loss of coherence catastrophe in Fourier transform ion cyclotron resonance mass spectrometry.
J Am Soc Mass Spectrom
; 20(2): 247-56, 2009 Feb.
Article
in En
| MEDLINE
| ID: mdl-19013078
ABSTRACT
The spontaneous loss of coherence catastrophe (SLCC) is a frequently observed, yet poorly studied, space-charge related effect in Fourier-transform ion cyclotron resonance mass spectrometry (FTICR-MS). This manuscript presents an application of the filter diagonalization method (FDM) in the analysis of this phenomenon. The temporal frequency behavior reproduced by frequency shift analysis using the FDM shows the complex nature of the SLCC, which can be explained by a combination of factors occurring concurrently, governed by electrostatics and ion packet trajectories inside the ICR cell.