Your browser doesn't support javascript.
loading
Homogeneous time-resolved fluoroimmunoassay for the measurement of midregional proadrenomedullin in plasma on the fully automated system B.R.A.H.M.S KRYPTOR.
Caruhel, Pascaline; Mazier, Christian; Kunde, Jan; Morgenthaler, Nils G; Darbouret, Bruno.
Affiliation
  • Caruhel P; Research and Development Department, CEZANNE SAS, Parc Scientifique Georges Besse, Nimes Cedex 1, France. pcaruhel@cezanne.fr
Clin Biochem ; 42(7-8): 725-8, 2009 May.
Article in En | MEDLINE | ID: mdl-19318039
ABSTRACT

OBJECTIVES:

Evaluate the first automated assay measuring a mid regional fragment of the proadrenomedullin (MR-proADM). DESIGN AND

METHODS:

B.R.A.H.M.S MR-proADM KRYPTOR is a homogeneous assay.

RESULTS:

In 144 healthy individuals, MR-proADM mean (SD) was measured at 0.37 nmol/L (0.09). High correlation with the manual assay, B.R.A.H.M.S MR-proADM LIA, was found on 281 samples from patients suffering from community acquired pneumonia (correlation coefficient 0.97).

CONCLUSIONS:

This fast assay can be used in low respiratory tract infections.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Protein Precursors / Fluoroimmunoassay / Adrenomedullin Limits: Adult / Female / Humans / Male / Middle aged Language: En Journal: Clin Biochem Year: 2009 Type: Article Affiliation country: France