Your browser doesn't support javascript.
loading
Analysis of early failure of the locking compression plate in osteoporotic proximal humerus fractures.
Micic, Ivan D; Kim, Kyung-Chun; Shin, Dong-Ju; Shin, Sang-Jin; Kim, Poong-Taek; Park, Il-Hyung; Jeon, In-Ho.
Affiliation
  • Micic ID; Department of Orthopaedic Surgery and Traumatology, Clinical Center, Faculty of Medicine, Nis, Serbia.
J Orthop Sci ; 14(5): 596-601, 2009 Sep.
Article in En | MEDLINE | ID: mdl-19802672
ABSTRACT

BACKGROUND:

Although there has been continuous evolution in the management of fracture fixation, treatment for osteoporotic proximal humerus fractures is still challenging to trauma surgeons. The purpose of this study was to report early failure of the locking compression plate (LCP) in the treatment of osteoporotic proximal humerus fracture and characterize the mode of failure.

METHODS:

Nine patients, older than 65 years, underwent internal fixation with the use of a locking compression plate and had early failure within 4 weeks postoperatively. According to Neer's classification, five were included in a two-part surgical neck fracture, three in a three-part fracture, and one in a four-part fracture.

RESULTS:

All failures occurred with back-out of the plate-screw construct, leading to varus displacement in eight patients and plate breakage in one. Revision surgery was performed in six patients using replating and tension band wiring with a bone graft, and three patients underwent hemiarthroplasty. The average UCLA score was 25 points for the hemiarthroplasty group and 30 points for the reconstruction group.

CONCLUSIONS:

Early postoperative failure of the LCP developed within 4 weeks with a presentation of en bloc back-out of the plate-screw construct and plate breakage. Possible risk factors included malreduction, loss of medial support, and negligence of tension band sutures on the tuberosities.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Shoulder Fractures / Bone Plates / Equipment Failure Analysis / Fracture Fixation, Internal Type of study: Risk_factors_studies Limits: Aged / Female / Humans / Male Language: En Journal: J Orthop Sci Journal subject: ORTOPEDIA Year: 2009 Type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Shoulder Fractures / Bone Plates / Equipment Failure Analysis / Fracture Fixation, Internal Type of study: Risk_factors_studies Limits: Aged / Female / Humans / Male Language: En Journal: J Orthop Sci Journal subject: ORTOPEDIA Year: 2009 Type: Article