Your browser doesn't support javascript.
loading
Radiation reduction in pediatric and adult congenital patients during cardiac catheterization.
Mauriello, Daniel A; Fetterly, Kenneth A; Lennon, Ryan J; Reeder, Guy S; Taggart, Nathaniel W; Hagler, Donald J; Cetta, Frank; Cabalka, Allison K.
Affiliation
  • Mauriello DA; Division of Pediatric Cardiology, Mayo Clinic, Rochester, Minnesota.
Catheter Cardiovasc Interv ; 84(5): 801-8, 2014 Nov 01.
Article in En | MEDLINE | ID: mdl-24782364
ABSTRACT

OBJECTIVES:

Our objective was to determine if technical changes combined with radiation safety initiatives reduced the radiation dose delivered to patients during congenital catheterization.

BACKGROUND:

Use of ionizing radiation is necessary during cardiac catheterization. Minimizing radiation dose, while maintaining clinically useful image quality, is an important safety issue. In our congenital heart center intentional practice changes, including technical changes and provider awareness initiatives, were implemented to decrease radiation dose.

METHODS:

Data were retrospectively collected for all procedures involving children and adults with congenital heart disease (CHD) undergoing catheterization over 45 months. Cases were divided into three categories including noninterventional (NI), simple intervention (SI), and complex intervention (CI). The change in dose was modeled as log of cumulative air kerma (Ka,r ). The change in Ka,r was evaluated for each procedural category as well as changes occurring as a function of age and weight.

RESULTS:

Considering all procedures (n = 1,082), Ka,r decreased by 61%. In the NI group (n = 481), Ka,r decreased by 71%. In the SI group (n = 424), Ka,r decreased by 74%. The Ka,r for the 10-17 year old group (n = 125) and those ≥18 years (n = 709) decreased 74 and 67%, respectively. The Ka,r decreased 72 and 66% for those 20-60 kg and ≥60 kg, respectively. Groups not showing significant change in Ka,r included CI, age ≤9 years, and weight ≤20 kg.

CONCLUSIONS:

Through technical changes and provider awareness initiatives, our institution dramatically reduced the radiation dose in the majority of pediatric and adult CHD patients undergoing cardiac catheterization.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Radiation Dosage / Radiation Protection / Cardiac Catheterization / Heart Defects, Congenital Type of study: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limits: Adult / Child / Child, preschool / Female / Humans / Infant / Male / Middle aged Language: En Journal: Catheter Cardiovasc Interv Journal subject: CARDIOLOGIA Year: 2014 Type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Radiation Dosage / Radiation Protection / Cardiac Catheterization / Heart Defects, Congenital Type of study: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limits: Adult / Child / Child, preschool / Female / Humans / Infant / Male / Middle aged Language: En Journal: Catheter Cardiovasc Interv Journal subject: CARDIOLOGIA Year: 2014 Type: Article