[The Development of Ammonia Sensor Based on Tunable Diode Laser Absorption Spectroscopy with Hollow Waveguide].
Guang Pu Xue Yu Guang Pu Fen Xi ; 36(8): 2669-73, 2016 Aug.
Article in Zh | MEDLINE | ID: mdl-30074727
Hollow waveguides(HWG)have recently emerged as a novel concept serving as an efficient optical waveguide and a highly miniaturized gas cell. Compared with conventional multi-pass gas cells, HWG gas cell has the advantages of facilitating gas exchanging because of its small size and fast responding speed. In this paper, we poposed an ammonia sensor based on tunable diode laser absorption spectroscopy(TDLAS) using HWG as the gas cell. The sensor employs wavelength modulation spectrum(WMS) with simultaneous detection of the second harmonic(2f) signal and the first hamonic(1f) signal. Normalization of the 2f signal by the 1f signal enables the sensor for calibration free measurement. The sensor performance is tested with gas standards and the result shows good linearity with correlation coefficient of 0.999 8, and the detection limit is 26 ppb with an integration time of 18 s. The sensor based on HWG gas cell is suitable for sensative and real-time monitoring ammonia in the air.
Search on Google
Collection: 01-internacional Database: MEDLINE Language: Zh Journal: Guang Pu Xue Yu Guang Pu Fen Xi Year: 2016 Type: Article