Your browser doesn't support javascript.
loading
Accuracy of cardiac functional parameters measured from gated radionuclide myocardial perfusion imaging in mice.
Hess, Annika; Nekolla, Stephan G; Meier, Martin; Bengel, Frank M; Thackeray, James T.
Affiliation
  • Hess A; Department of Nuclear Medicine, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
  • Nekolla SG; Department of Nuclear Medicine, Technical University of Munich, Munich, Germany.
  • Meier M; Imaging Center of the Institute of Laboratory Animal Sciences, Hannover Medical School, Hannover, Germany.
  • Bengel FM; Department of Nuclear Medicine, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany.
  • Thackeray JT; Department of Nuclear Medicine, Hannover Medical School, Carl-Neuberg-Str. 1, 30625, Hannover, Germany. Thackeray.James@mh-hannover.de.
J Nucl Cardiol ; 27(4): 1317-1327, 2020 08.
Article in En | MEDLINE | ID: mdl-31044402
BACKGROUND: Quantitative cardiac contractile function assessment is the primary indicator of disease progression and therapeutic efficacy in small animals. Operator dependency is a major challenge with commonly used echocardiography. Simultaneous assessment of cardiac perfusion and function in nuclear scans would reduce burden on the animal and facilitate longitudinal studies. We evaluated the accuracy of contractile function measurements obtained from electrocardiogram-gated nuclear perfusion imaging compared with anatomic imaging. METHODS AND RESULTS: In healthy C57Bl/6N mice (n = 11), 99mTc-sestamibi SPECT and 13N-ammonia PET underestimated left ventricular volumes (23 to 28%, P = 0.02) compared to matched anatomic images, though ejection fraction (LVEF) was comparable (%, SPECT: 73 ± 8 vs CMR: 72 ± 6, P = 0.1). At 1 week after myocardial infarction (n = 13), LV volumes were significantly lower in perfusion images compared to CMR and contrast CT (P = 0.003), and LVEF was modestly overestimated (%, SPECT: 37 ± 8, vs CMR: 27 ± 7, P = 0.003). Nuclear images exhibited good intra- and inter-reader agreement. Perfusion SPECT accurately calculated infarct size compared to histology (r = 0.95, P < 0.001). CONCLUSIONS: Cardiac function can be calculated by gated nuclear perfusion imaging in healthy mice. After infarction, perfusion imaging overestimates LVEF, which should be considered for comparison to other modalities. Combined functional and infarct size analysis may optimize imaging protocols and reduce anaesthesia duration for longitudinal studies.
Subject(s)
Key words

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Cardiac-Gated Single-Photon Emission Computer-Assisted Tomography / Myocardial Perfusion Imaging / Myocardial Contraction / Myocardial Infarction Type of study: Guideline / Observational_studies Limits: Animals Language: En Journal: J Nucl Cardiol Journal subject: CARDIOLOGIA Year: 2020 Type: Article Affiliation country: Germany

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Cardiac-Gated Single-Photon Emission Computer-Assisted Tomography / Myocardial Perfusion Imaging / Myocardial Contraction / Myocardial Infarction Type of study: Guideline / Observational_studies Limits: Animals Language: En Journal: J Nucl Cardiol Journal subject: CARDIOLOGIA Year: 2020 Type: Article Affiliation country: Germany