Ultrasound-guided deployment of ProGlide™ device in transfemoral transcatheter aortic valve implantation and risk reduction of vascular complications: A propensity-matched cohort study.
Catheter Cardiovasc Interv
; 103(5): 771-781, 2024 Apr.
Article
in En
| MEDLINE
| ID: mdl-38451155
ABSTRACT
BACKGROUND:
ProGlide is a percutaneous suture-mediated closure device used in arterial and venous closure following percutaneous intervention. Risk of vascular complications from use, particularly related to failure in hemostasis, or acute vessel closure, remains significant and often related to improper suture deployment. We describe a technique of ultrasound-guided ProGlide deployment in transfemoral transcatheter aortic valve implantation (TF-TAVI).AIMS:
The aim of this study is to assess vascular outcomes for ultrasound-guided deployment of ProGlide vascular closure devices in patients undergoing TF-TAVI.METHODS:
We collected relevant clinical data of patients undergoing TAVI in a large volume centre. PRIMARYOUTCOME:
main access Valve Academic Research Consortium 3 (VARC-3) major vascular complication. SECONDARYOUTCOME:
any major/minor VARC-3 vascular complication, its type (bleed or ischemia), and treatment required (medical, percutaneous, or surgical). We performed inverse weighting propensity score analysis to compare the population undergoing ultrasound-guided versus conventional ProGlide deployment for main TAVI access. Ultrasound technique for ProGlide insertion was performed as described below.RESULTS:
Five hundred and seventeen patients undergoing TF-TAVI were included. PRIMARYOUTCOME:
In 126 (ultrasound-guided) and 391 (conventional ProGlide insertion), 0% versus 1.8% (p < 0.001) had a major VARC-3 vascular complication, respectively. SECONDARYOUTCOME:
0.8% (one minor VARC-3 bleed) vs 4.1% (13 bleeds and three occlusions) had any VARC-3 vascular complication (major and minor) (p < 0.001). Surgical treatment of vascular complication was required in 0.8% versus 1.3% (p = NS).CONCLUSIONS:
Ultrasound-guided deployment of ProGlide for vascular closure reduced the risk of major vascular complications in a large population undergoing TAVI.Key words
Full text:
1
Collection:
01-internacional
Database:
MEDLINE
Main subject:
Aortic Valve Stenosis
/
Transcatheter Aortic Valve Replacement
/
Vascular Closure Devices
Limits:
Humans
Language:
En
Journal:
Catheter Cardiovasc Interv
Journal subject:
CARDIOLOGIA
Year:
2024
Type: