Your browser doesn't support javascript.
loading
Characterization of Microwave Generator Energy and Ablation Volumes following Transarterial Embolization in an In Vivo Porcine Liver Model.
Sparks, Hiro; Rink, Johann S; Ramakrishnan, Abinaya; Sung, Kyunghun; Ni, Jason; Lu, David S K; Raman, Steven S; Lee, Edward W; Chiang, Jason.
Affiliation
  • Sparks H; Department of Radiology, Ronald Reagan University of California, Los Angeles Medical Center, Los Angeles, California. Electronic address: HDSparks@mednet.ucla.edu.
  • Rink JS; Department for Radiology and Nuclear Medicine, Mannheim University Medical Centre, Mannheim, Germany.
  • Ramakrishnan A; Department of Radiology, Ronald Reagan University of California, Los Angeles Medical Center, Los Angeles, California.
  • Sung K; Department of Radiology, Ronald Reagan University of California, Los Angeles Medical Center, Los Angeles, California.
  • Ni J; Department of Radiology, Ronald Reagan University of California, Los Angeles Medical Center, Los Angeles, California.
  • Lu DSK; Department of Radiology, Ronald Reagan University of California, Los Angeles Medical Center, Los Angeles, California.
  • Raman SS; Department of Radiology, Ronald Reagan University of California, Los Angeles Medical Center, Los Angeles, California.
  • Lee EW; Department of Radiology, Ronald Reagan University of California, Los Angeles Medical Center, Los Angeles, California.
  • Chiang J; Department of Radiology, Ronald Reagan University of California, Los Angeles Medical Center, Los Angeles, California.
J Vasc Interv Radiol ; 35(8): 1227-1233, 2024 Aug.
Article in En | MEDLINE | ID: mdl-38754759
ABSTRACT

PURPOSE:

To characterize the relationship between ablation zone volume (AZV) and microwave ablation (MWA) energy in an in vivo porcine liver model following arterial embolization. MATERIALS AND

METHODS:

With Institutional Animal Care and Use Committee (IACUC) approval, 11 female swine underwent either right (n = 5) or left (n = 6) hepatic artery embolization under fluoroscopic guidance. Subsequently, ultrasound (US)-guided MWA was performed in each liver segment (left lateral, left medial, right medial, and right lateral) at either 30 W (n = 4 lobes), 60 W (n = 4), 65 W (n = 20), 90 W (n = 8), 120 W (n = 4), or 140 W (n = 4) continuously for 5 minutes. Postprocedural volumetric segmentation was performed on standardized multiphase T1 magnetic resonance (MR) imaging sequences.

RESULTS:

Mean AZVs in embolized lobes (15.8 mL ± SD 10.6) were significantly larger than those in nonembolized lobes (11.2 mL ± SD 6.5, P < .01). MWA energy demonstrated significant positive linear correlation with both embolized (R2 = 0.66, P < .01) and nonembolized (R2 = 0.64, P < .01) lobes. The slope of the linear models corresponded to a 0.95 mL/kJ (SD ± 0.16) and 0.54 mL/kJ (SD ± 0.09) increase in ablation volume per applied kilojoule of energy (E) in embolized and nonembolized lobes, respectively. In the multivariate model, embolization status significantly modified the relationship between E and AZV as described by the following interaction term 0.42∗E∗(embolization status) (P = .031).

CONCLUSIONS:

Linear models demonstrated a near 1.8-fold increase in ratio of AZV per unit E, R(AZVE), when applied to embolized lobes relative to nonembolized lobes. Absolute AZV differences between embolized and nonembolized lobes were greater at higher-power MWA.
Subject(s)

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Models, Animal / Embolization, Therapeutic / Hepatic Artery / Liver / Microwaves Limits: Animals Language: En Journal: J Vasc Interv Radiol Journal subject: ANGIOLOGIA / RADIOLOGIA Year: 2024 Type: Article

Full text: 1 Collection: 01-internacional Database: MEDLINE Main subject: Models, Animal / Embolization, Therapeutic / Hepatic Artery / Liver / Microwaves Limits: Animals Language: En Journal: J Vasc Interv Radiol Journal subject: ANGIOLOGIA / RADIOLOGIA Year: 2024 Type: Article