Your browser doesn't support javascript.
loading
Method for evaluating optical characteristics of endoscopes for recording fluorescence-related cardiac electrical activity.
Sau, Soma; Bassen, Howard I; Krauthamer, Victor.
Afiliación
  • Sau S; Department of Biomedical Engineering, Duke University, Durham, North Carolina 27708, USA.
Appl Opt ; 41(31): 6725-30, 2002 Nov 01.
Article en En | MEDLINE | ID: mdl-12412663
ABSTRACT
Nondestructive methods were used to evaluate marketed fiber-optic endoscopes (intended for simple viewing) for fluorescence recording. Our application is for optical recording from the heart. For one angioscope, we measured a focal length of 0.33 mm, a field of view of 45 degrees, an aperture of 0.26 mm, and an efficiency of 43%. We calculated that the angioscope would give a signal-to-noise ratio of 1.0 for a cardiac action potential, if its field of view were divided into a nine-pixel array (for safe continuous illumination). Our methods are useful in designing and evaluating fluorescence fiber-optic systems with superior signal quality and spatial resolution.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Endoscopios / Electrofisiología / Corazón Idioma: En Revista: Appl Opt Año: 2002 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Endoscopios / Electrofisiología / Corazón Idioma: En Revista: Appl Opt Año: 2002 Tipo del documento: Article País de afiliación: Estados Unidos