Determination of kurarinone in rat plasma by UPLC-MS/MS.
Zhang, Wei-min; Li, Rui-fang; Qiu, Jian-fei; Zhang, Zhi-yin; Wang, Hong-bo; Bian, Lu; Lei, Jia-hui.
Afiliación
  • Zhang WM; The First Affiliated Hospital of Henan University of Science and Technology, Luoyang, Henan 471003, PR China. Electronic address: lyzhangweimin@163.com.
  • Li RF; Medical College of Henan University of Science and Technology, Luoyang, Henan 471003, PR China.
  • Qiu JF; Medical College of Henan University of Science and Technology, Luoyang, Henan 471003, PR China.
  • Zhang ZY; Medical College of Henan University of Science and Technology, Luoyang, Henan 471003, PR China.
  • Wang HB; Medical College of Henan University of Science and Technology, Luoyang, Henan 471003, PR China.
  • Bian L; Medical College of Henan University of Science and Technology, Luoyang, Henan 471003, PR China.
  • Lei JH; Medical College of Henan University of Science and Technology, Luoyang, Henan 471003, PR China.
Article en En | MEDLINE | ID: mdl-25703950
ABSTRACT
A sensitive and rapid ultra performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS) method was developed to determine kurarinone in rat plasma using chlorzoxazone as the internal standard (IS). Sample preparation was accomplished through a liquid-liquid extraction procedure with ethyl acetate to 0.2 mL plasma sample. The analyte and IS were separated on an Acquity UPLC BEH C18 column (2.1 mm × 50 mm, 1.7 µm) with the mobile phase of acetonitrile and 1% formic acid in water with gradient elution at a flow rate of 0.40 mL/min. The detection was performed on a triple quadrupole tandem mass spectrometer equipped with electrospray ionization (ESI) by multiple reactions monitoring (MRM) of the transitions at m/z 437.0→301.2 for kurarinone and m/z 168.1→132.1 for IS. The linearity of this method was found to be within the concentration range of 20-2000 ng/mL with a lower limit of quantification of 20 ng/mL. Only 3.0 min was needed for an analytical run. The matrix effect was 94.7-107.2% for kurarinone. The intra- and inter-day precision (RSD%) were less than 8.2% and accuracy (RE%) was within ±9.0%. The recovery ranged from 77.3% to 85.6%. Kurarinone was sufficiently stable under all relevant analytical conditions. The method was also successfully applied to the pharmacokinetic study of kurarinone in rats.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Flavonoides / Cromatografía Líquida de Alta Presión / Espectrometría de Masas en Tándem Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Animals Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2015 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Flavonoides / Cromatografía Líquida de Alta Presión / Espectrometría de Masas en Tándem Tipo de estudio: Diagnostic_studies / Prognostic_studies Límite: Animals Idioma: En Revista: J Chromatogr B Analyt Technol Biomed Life Sci Asunto de la revista: ENGENHARIA BIOMEDICA Año: 2015 Tipo del documento: Article