Marginal and internal fit of pressed ceramic crowns made from conventional and computer-aided design and computer-aided manufacturing wax patterns: An in vitro comparison.
Shamseddine, Loubna; Mortada, Rola; Rifai, Khaldoun; Chidiac, Jose Johann.
Afiliación
  • Shamseddine L; Assistant Professor, Prosthodontics Department, Lebanese University School of Dentistry, Hadath, Lebanon. Electronic address: drloubna1@hotmail.com.
  • Mortada R; Clinical instructor, Prosthodontics Department, Lebanese University School of Dentistry, Hadath, Lebanon.
  • Rifai K; Professor, Prosthodontics Department, Lebanese University School of Dentistry, Hadath, Lebanon.
  • Chidiac JJ; Professor, Prosthodontics Department, Lebanese University, School of Dentistry, Hadath, Lebanon.
J Prosthet Dent ; 116(2): 242-8, 2016 Aug.
Article en En | MEDLINE | ID: mdl-26948080
ABSTRACT
STATEMENT OF

PROBLEM:

To our knowledge, no studies have evaluated the influence of the wax fabrication technique on the internal adaptation and marginal accuracy of pressed ceramic single crowns.

PURPOSE:

The purpose of this in vitro study was to compare the marginal and internal fit of a lithium disilicate crown fabricated either from conventional waxing or from a subtractive milled wax pattern. MATERIAL AND

METHODS:

Ten silicone impressions were made for a maxillary canine prepared to receive a complete crown. Two pressed lithium disilicate glass ceramic copings were made on the die poured from each impression. The first was from a conventional wax pattern and the second from milled wax blocks. The replica technique was used to measure the fit by scanning electron microscope at ×80 magnification. Collected data were analyzed with the paired Student t test for the marginal and wall fit. For the occlusal fit, the difference in scores did not follow a normal distribution, and the Wilcoxon signed ranks test was used (α=.05).

RESULTS:

The mean values +SD were 170.3 ±50.7 µm for the marginal fit and 130.3 ±21.6 µm for the internal fit with the conventional technique. For the computer-aided design and computer-aided manufacturing (CAD-CAM) technique, they were 105.1 ±39.6 µm and 98.1 ±26.1 µm. Statistical significance was found for the marginal and internal fit (P<.05). The occlusal fit median and interval interquartile scores were 305.0 µm (269.9 to 334.4) for conventional fabrication and 199.0 µm (141.5 to 269.9) for subtractive manufacturing. The differences were not statistically significant (P>.05).

CONCLUSIONS:

The subtractive CAD-CAM waxing technique resulted in the improved fit of a pressed lithium disilicate crown by decreasing the marginal discrepancies and internal adaptation.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Diseño de Prótesis Dental / Diseño Asistido por Computadora / Adaptación Marginal Dental / Coronas Límite: Humans Idioma: En Revista: J Prosthet Dent Año: 2016 Tipo del documento: Article