The robustness of T2 value as a trabecular structural index at multiple spatial resolutions of 7 Tesla MRI.
Lee, D K; Song, Y K; Park, B W; Cho, H P; Yeom, J S; Cho, G; Cho, H.
Afiliación
  • Lee DK; Department of Biomedical Engineering, Ulsan National Institute of Science and Technology, Ulsan, South Korea.
  • Song YK; Korea Basic Science Institute, Ochang, South Korea.
  • Park BW; Department of Biomedical Engineering, Ulsan National Institute of Science and Technology, Ulsan, South Korea.
  • Cho HP; Department of Biomedical Engineering, Ulsan National Institute of Science and Technology, Ulsan, South Korea.
  • Yeom JS; Department of Orthopedic Surgery, Seoul National University, Seoul, South Korea.
  • Cho G; Korea Basic Science Institute, Ochang, South Korea.
  • Cho H; Department of Biomedical Engineering, Ulsan National Institute of Science and Technology, Ulsan, South Korea.
Magn Reson Med ; 80(5): 1949-1961, 2018 11.
Article en En | MEDLINE | ID: mdl-29656389
ABSTRACT

PURPOSE:

To evaluate the robustness of MR transverse relaxation times of trabecular bone from spin-echo and gradient-echo acquisitions at multiple spatial resolutions of 7 T.

METHODS:

The effects of MRI resolutions to T2 and T2* of trabecular bone were numerically evaluated by Monte Carlo simulations. T2 , T2*, and trabecular structural indices from multislice multi-echo and UTE acquisitions were measured in defatted human distal femoral condyles on a 7 T scanner. Reference structural indices were extracted from high-resolution microcomputed tomography images. For bovine knee trabecular samples with intact bone marrow, T2 and T2* were measured by degrading spatial resolutions on a 7 T system.

RESULTS:

In the defatted trabecular experiment, both T2 and T2* values showed strong ( |r| > 0.80) correlations with trabecular spacing and number, at a high spatial resolution of 125 µm3 . The correlations for MR image-segmentation-derived structural indices were significantly degraded ( |r| < 0.50) at spatial resolutions of 250 and 500 µm3 . The correlations for T2* rapidly dropped ( |r| < 0.50) at a spatial resolution of 500 µm3 , whereas those for T2 remained consistently high ( |r| > 0.85). In the bovine trabecular experiments with intact marrow, low-resolution (approximately 1 mm3 , 2 minutes) T2 values did not shorten ( |r| > 0.95 with respect to approximately 0.4 mm3 , 11 minutes) and maintained consistent correlations ( |r| > 0.70) with respect to trabecular spacing (turbo spin echo, 22.5 minutes).

CONCLUSION:

T2 measurements of trabeculae at 7 T are robust with degrading spatial resolution and may be preferable in assessing trabecular spacing index with reduced scan time, when high-resolution 3D micro-MRI is difficult to obtain.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Imagen por Resonancia Magnética / Hueso Esponjoso Tipo de estudio: Health_economic_evaluation Límite: Animals Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2018 Tipo del documento: Article País de afiliación: Corea del Sur

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Procesamiento de Imagen Asistido por Computador / Imagen por Resonancia Magnética / Hueso Esponjoso Tipo de estudio: Health_economic_evaluation Límite: Animals Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2018 Tipo del documento: Article País de afiliación: Corea del Sur