Monte Carlo simulation study on the cause of the formation of the efficiency curve at low energies in germanium detectors.
Appl Radiat Isot
; 192: 110571, 2023 Feb.
Article
en En
| MEDLINE
| ID: mdl-36463617
Detector efficiency is a measure of the detectors' ability to detect radiation. It is known that the efficiency values, which are important and affect the activity calculation in germanium detectors, decrease rapidly at low energies. This study focuses on the reasons for this handicap in the low energy region of the efficiency curve. The Monte Carlo simulation was carried out with four setups: germanium only, germanium with a dead layer, germanium with aluminum holders, and finally germanium with a dead layer and aluminum holders. The effect of each setup on the efficiency curve was observed. As a result, it was seen that the main cause of the handicap was the dead layer. For this reason, it was concluded that the current value of the dead layer, which is known to change over time, should be taken into account in detector calibrations or characterizations.
Palabras clave
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Germanio
Idioma:
En
Revista:
Appl Radiat Isot
Asunto de la revista:
MEDICINA NUCLEAR
/
SAUDE AMBIENTAL
Año:
2023
Tipo del documento:
Article